User Performance Evaluation of an Investigational Blood Glucose Monitoring System

The purpose of this study is to demonstrate that untrained subjects with diabetes can use an investigational Blood Glucose Monitoring System (BGMS) with capillary blood obtained from fingerstick and from Alternative Site (AST) Palm .

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Bayer HealthCare LLC, Diabetes Care

Mishawaka, Indiana, 46544, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Males and females, 18 years of age and above
  • Diagnosed as having type 1 or type 2 diabetes
  • Currently self-testing blood glucose at home
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion criteria

  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Employee of competitive medical device company,immediate family member of such, or living within the household of such
  • Pregnancy

Treatment and study plan

Ninja 2 Investigational Blood Glucose Monitoring System

Device

The Ninja 2 meter is a Bayer investigational meter that uses an investigational sensor.

Primary outcomes

  1. Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL(<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Meter (BGM). BGM results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results <75mg/dL) or within +/- 20%(for reference BG results >=75mg/dL) of the reference method results (YSI capillary plasma).

Secondary outcomes

  1. Number of Alternative Site (AST) Palm Blood Glucose (BG) Results Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Untrained subjects with diabetes self-test subject Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGM meter results are compared with capillary plasma BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results are used to calculate the number of BG results within +/- 15mg/dL (for reference BG results <75mg/dL) or +/- 20% (for reference BG results >=75mg/dL) of the YSI capillary plasma reference method results.

Sponsors and collaborators

Lead sponsor

Ascensia Diabetes Care

Industry

Registry information

Official study title

Ninja 2 User Performance Evaluation

Important dates

Study start
2011
Primary completion
2011
Study completion
2011
First posted
Aug 5, 2011
Registry last updated
Feb 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.