Usefulness of Non-invasive Pulse Co-oximetry Haemaglobin Measurements in Critically Ill Black Patients

To answer the question whether a non-invasive haemaglobin measurement is clinically useful, reliable and accurate as compared to taking a blood sample and checking the haemaglobin level at the laboratory or in a blood gas analyser. This study will take place in a multi-disciplinary ICU of critically ill patients.

Completed

Looking for future studies?

Notify Me

Key information

Age range

1 month–100 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Chris Hani Baragwanath Hospital

Soweto, Gauteng, South Africa

About this study

A presenting sample of 150 patients requiring admission to ICU will be enrolled. These will include paediatric, trauma, adult medical and adult surgical patients. Informed consent will be obtained. Baseline demographic data, vital signs, Massey pigmentation score, and severity of illness scores will be calculated, as well as finger deformity, if present, nail polish or acrylics, smoking habits, finger diameter of finger measured, comorbidities and medications. Patients will be admitted in the usual way, and admission bloods will be sent to the laboratory as per usual protocol. Enrolled patients will in addition have their Haemaglobin and Plethysmography Index (measure of perfusion) measured non-invasively using the Masimo Pronto-7 handheld device. Note of concurrent medications, blood products and vital signs will be recorded at each measurement. Each patient will have measurements done 8 hourly (between 1-5 measurements per patient). Concurrent arterial blood gas samples will be taken in a heparinised syringe and performed on ABL radiometer blood gas analyser.. An additional EDTA blood sample shall be taken at each Pronto measuring point which will be measured at the laboratory on a Sysmex cell counter. Analysis of data will assess precision and accuracy, trend accuracy, and effect of pigmentation, vasopressors and other medication on the results of the non-invasive co-oximetry estimation of haemaglobin.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • All patients with a condition that requires admission to ICU shall be considered to be eligible.
  • Age 1month to 100years old

Exclusion criteria

  • Patients under 1 month old
  • Patients with an unrecordable blood pressure or body temperature <34 degrees.

Treatment and study plan

non invasive co-oximetry haemaglobin measurement

Other

Primary outcomes

  1. Accuracy and precision of non invasive haemaglobin measurement.

    Time frame: during ICU admission

    Accuracy and precision of non invasive hemoglobin (Hb) measurement to invasive co-oximetry (blood gas analyser), and laboratory haemaglobin measurement.

    For comparative purposes in our population (dark skinned patients, low Hb levels and during active transfusion) our outcome aim for Hb accuracy will be based on what Masimo has found previously in 11 335 comparisons. These are:

    • 0.99g/dl at 1SD
    • Hb between 6g/dl and 12g/dl : 95% of readings within 2g/dl of laboratory value
    • Hb between 12g/dl and 18g/dl : 95% of readings within 2g/dl of laboratory value

    Precision shall be described as a co-efficient of variation

Secondary outcomes

  1. Time to result

    Time frame: during ICU admission

    Time to result of non invasive hemoglobin (Hb) measurement compared to invasive co-oximetry (blood gas analyser), and laboratory haemaglobin measurement (Sysmex cell counter)

  2. Cost of test

    Time frame: during ICU admission

    To compare the non invasive hemoglobin (Hb) measurement to invasive co-oximetry (blood gas analyser), and laboratory haemaglobin measurement (Sysmex cell counter) with regards to cost.

  3. Effect of skin pigmentation on result

    Time frame: during ICU admission

    Effect of skin pigmentation on result of non invasive hemoglobin (Hb) measurement compared to to invasive co-oximetry (blood gas analyser), and laboratory haemaglobin measurement (Sysmex cell counter)

  4. Effect of patients clinical state on test results.

    Time frame: during ICU admission

    Effect of patients clinical state on test results of non invasive hemoglobin (Hb) measurement compared to invasive co-oximetry (blood gas analyser), and laboratory haemaglobin measurement (Sysmex cell counter).

    The effect of patients clinical state (temp, MAP, pH, Hb level, plethysmography index (PI), severity of illness score, presence of active transfusion, presence of active bleeding, use of pressors, use of other blood products) on test results (accuracy and precision) shall be evaluated.

Sponsors and collaborators

Lead sponsor

Chris Hani Baragwanath Academic Hospital

Other

Collaborators

  • Masimo Corporation

Registry information

Important dates

Study start
2014
Primary completion
2016
Study completion
2016
First posted
Jan 14, 2014
Registry last updated
Aug 16, 2016