NCT Number: NCT01841255

UMOX™: A New Device for Preoxygenation

Background - Preoxygenation is performed before general anesthesia to increase the oxygen reserves of the body and prolong the safe period of apnea without hypoxia. However, it is often impossible to obtain a perfect face mask fit, and optimal end-tidal fractional oxygen concentration (EtO2) are not attained. The investigators designed a new oropharyngeal preoxygenation device, UMOX™, to avoid leaks during preoxygenation. In this study, the investigators compared the efficacy of UMOX™ with that of a conventional face mask.

Methods - In 50 healthy volunteers breathing 100% oxygen from a circle absorber system for a 5 minute-period, the investigators compared preoxygenation using the 1) a conventional mask; 2) UMOX™ without any instruction; 3) UMOX™ with instruction on mouth breathing; and 4) UMOX™ with a nose clip, in random order. Each subject underwent all steps with a 10-minute rest period of room air breathing between each step. Inspired and end-tidal respiratory gases were measured every 10 seconds. EtO2 variation was analyzed using Anova.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–65 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Maisonneuve-Rosemont Hospital

Montreal, Quebec, H1T 2M4, Canada

About this study

This study was as a single center, crossover, randomized trial. Following approval from research and ethics committee from our hospital and written consent from healthy subjects recruited, 50 consenting volunteers were enrolled over a three-month period, from: September 29, 2008 to November 5, 2008.

Subjects enrolled were of both genders, smokers and non-smokers, American Society of Anesthesiologists (ASA) physical status I or II, 18 to 65 years old. Nonrecruitment depended on the presence of any craniofacial abnormality, facial hair (beard or moustache) or absence of teeth. Well-controlled non-severe asthma was not a criterion for exclusion.

The trial was conducted in an operating room, using the Dräger Fabius GS (Dräger, Canada), anesthesia machine in a usual set up with the operating table. The investigators used a, patient circuit system consists of 1.5-L lime soda absorber connected to disposable corrugated tubing and a 2-L breathing bag. Gases were analyzed through a Marquette Medical Systems™ SAM module (GE Healthcare), volumes monitored with a GE Respirator Mechanics-M (GE Healthcare) and SpO2 measured with a Nellcor™ (Covidien) pulse oximetry positioned on a finger. Preoxygenation procedures were explained to the subjects and ample time was allowed so they could become familiar with the mask and mouth piece. Set up for the measure was conducted by the same experimenter. Subjects laid supine and preoxygenation was performed with fresh gas flow: 100% oxygen (O2) 10L/minute, administered trough facemask or mouth piece sequentially in a random order. Randomization was established with a random assignment table (ww.randomization.com). The envelope was open as the subject entered the operating room (OR). At the beginning and between each step, the anesthesia circuit was flushed with 100% (O2). Each subject went through all steps in random order, each requiring a 5-min period of: 1) tidal breathing using the facemask (TVFM); 2) tidal volume breathing with the UMOXTM mouthpiece (TVMP); 3) tidal volume breathing with the UMOXTM mouthpiece and verbal instructions to favor mouth breathing (TVMPMB); (4) tidal volume breathing with the UMOXTM mouthpiece and a nose clip (TVMPNC). The experimenter conducting the steps instructed participants to create a seal with their lips around the mouthpiece and readjusted the facemask when the subject signalled a discernable leak. Each step was interrupted by 10 minutes of breathing normal room air to allow for return to original respiratory status. Continuous recording of experimental measures, displayed on the anesthesia monitoring system, was done on digital films (Canon PowershotS5 IS™). For the purpose of our trial, 10 seconds intervals were considered adequate for following preoxygenation trends. End-tidal O2 was measured continuously for 5 min. We have also evaluated the level of satisfaction of the subject with each technique using a 1-4 scale.Participants were asked to asses their comfort by showing the associated number with their fingers. Finally we described the performance of the facemask ventilation technique as reported by the experimenter on a scale of 1 to 4.Primary outcome was the end-tidal O2 at the end of the 5-minutes preoxygenation period. Secondary outcomes included endtidal O2 as a function of time, the number of subjects reaching at least 87% EtO2, subject satisfaction and ease of ventilation.

Based on the few studies available on different preoxygenation technques, the end-tidal oxygen fraction standard deviation can range from 5% to about 20% depending on the device used for preoxygenation3,4,14,15. Assuming a 12% value for standard deviation, sample size calculations were made for Type I and Type II errors of 5% and 20%, respectively (power of 80%). From these numbers and taking account that we would rely on multiple comparisons, the power analysis suggested that 50 subjects would allow us to find a difference between groups around 9%. The 9nvestigators analyzed the EtO2 final value with one-way ANOVA followed by the Tukey's Multiple Comparisons test. Percentage of participants reaching at least 87% EtO2 are described with survival curves compared with Kaplan-Meier log rank test. Tolerance is presented using Kruskal-Wallis with Dunn's Multiple Comparisons test.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Both genders
  • smokers and non-smokers
  • American Society of Anesthesiologists (ASA) physical status I or II
  • 18 to 65 years old.

Exclusion criteria

  • Presence of any craniofacial abnormality
  • facial hair (beard or moustache) or absence of teeth.
  • Well-controlled non-severe asthma was not a criterion for exclusion.

Treatment and study plan

Mouthpiece (Umox)

Device

Device used to perform preoxygenation in the operating room context

Other names: Umox

Facemask

Device

Primary outcomes

  1. The primary outcome measurement was the evolution of endtidal oxygen fraction during a 5 minute preoxygenation procedure, measurements were done continuously, maximum values were recorded

    Time frame: The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

    The goal of this study is to compare the efficacy of preoxygenation using the conventional mask with the mouthpiece of the UMOX™ oropharyngeal device. The primary outcome measurement was endtidal oxygen fraction. The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers. The measures were not part of a surgical procedure, participants were volunteers participating in this experimental design. No follow-up step was included in this protocol.

Secondary outcomes

  1. The investigators also evaluated the respective tolerability of the two preoxygenation instruments in healthy volunteers with a short questionnaire

    Time frame: The four steps were done sequentially, each step consisted of a 5 minute recording period under specific conditions, followed by a 10 minute pause before the next step

Sponsors and collaborators

Lead sponsor

Maisonneuve-Rosemont Hospital

Other

Registry information

Important dates

Study start
2008
Primary completion
2008
Study completion
2008
First posted
Apr 26, 2013
Registry last updated
Apr 26, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)