NCT Number: NCT00588484
Develop a standard technique and demonstrate adequate precision for measurement of intima-media thickness (IMT) for routine clinical use.
Looking for future studies?
Notify Me35 year–65 year
All sexes
Observational
Rochester, Minnesota, 55905, United States
IMT is measured using software to analyze ECG-gated US image sequences acquired using a standard protocol. Two sonographers acquire images and measure IMT for 100 vessels. IMT technique standard deviation (SD) is determined by assuming normally distributed measurements, and dividing the SD of the 100 (paired) IMT differences by sqrt(2). This SD will be taken as measures of precision of the IMT technique.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: Four years
Time frame: Two years
Mayo Clinic
Other
Determination of the Precision and Clinical Utility of Ultrasound-based Intima-Media Thickness Measurements
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.