U-TruSignal SpO2 Testing in Neonates

The purpose of the study is to demonstrate proper function of the U-TruSignal device via clinical performance testing in a neonatal human subject population under standard clinical conditions. A study with human subjects will provide the needed clinical evidence for assessing the accuracy of the pulse oximeter as recommended by the FDA Guidance Document (Pulse Oximeters - Premarket Notification Submissions [50(k)s]: Guidance for Industry and Food and Drug Administration Staff.)

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

Up to 29 day

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Tampere University Hospital (TAYS), Tampere, Finland

Loading trial locations.

About this study

The purpose of this study is to collect blood samples from neonates under normal clinical conditions to ensure proper function (clinical performance) with the U-TruSignal device.

The study will include both the TruSignal AllFit Sensor and TruSignal Sensitive Sensor.

After providing consent, no preparation beyond the investigational site's standard of care is required before study procedures begin. At the start of the procedure, one to two sensors, upon the discretion of the investigator, shall be applied to the subject.

The duration of the subject's participation in the study is dependent upon the scheduled routine arterial blood draw and obtaining the maximum of 6 data pairs (3 data pairs per sensor). The Investigator or designated study staff will collect of SpO2 and pulse rate for each study subject. Laboratory results from arterial blood draw will be collected from the hospital's medical records. The description, severity, and device relatedness of any AE or SAE during the study will be recorded.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Parents or legally authorized representative (LAR) can understand and provide written informed consent; AND
  • Subjects are < 29 days old, and requiring arterial blood samples per the site's standard of care.

Exclusion criteria

  • Neonates with injuries, deformities or abnormalities may prevent proper application of the sensor;
  • Neonates with inadequate heart pump function affecting blood circulation caused by identified cardiac or cardiovascular conditions such as patent ductus arteriosus, persistent pulmonary hypertension, septal defects, or congenital heart disease;
  • Neonates with mean arterial blood pressure < 20mmHg;
  • Neonates with congenital diaphragmatic hernia; OR
  • Neonates under High frequency ventilation therapy.

Treatment and study plan

SpO2 Sensor application & blood draw

Device

At the start of the procedure, one to two commercial sensors shall be applied to the subject. Readings will be allowed to stabilize. After at least 10 minutes of data collection, arterial blood will be collected and analyzed blood per hospital's standard procedure. Data collection will continue post blood draw for two minutes.

Primary outcomes

  1. Accuracy Root Mean Square (ARMS) Per Data Pair

    Time frame: 30 minutes per data pair

    Compare observations of Peripheral Capillary Oxygen Saturation (SpO2- measured by the test pulse oximeter system) and Oxygen Saturation of Arterial Blood (SaO2) values (measured by CO-Oximeter analysis of simultaneous drawn arterial blood).

Secondary outcomes

  1. Continuous SpO2 Measurements During Data Collection Interval

    Time frame: 30 minutes per data pair

    The secondary objective of the study was to demonstrate U-TruSignal collects substantially continuous SpO2 measurements during a data collection interval. All time the sensors were worn and data collected across participants were reviewed for invalid data within each data pair. For the data pairs that had moments of invalid data, a calculation was conducted (invalid data/complete data for data pair) to establish percentage of invalid data per sensor. For each sensor type the percentage of invalid data (%) was calculated. Sensor readings were grouped together across participants and are being reported as the total number of invalid readings over the total number of readings for all participants

Other outcomes

  1. Type and Number of AEs, SAEs, and Device Issues

    Time frame: 30 minutes per data pair

    Collect safety information, including type and number of AE s, SAEs, and device issues.

Sponsors and collaborators

Lead sponsor

GE Healthcare

Industry

Registry information

Important dates

Study start
2017
Primary completion
2019
Study completion
2019
First posted
Dec 26, 2017
Registry last updated
Dec 4, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.