Skip to main content
OpenTrials
Completed

NCT Number: NCT01097772

TriVascular European Union (EU) Abdominal Stent Graft Trial

A Clinical Study to Evaluate the Safety and Performance of the TriVascular Ovation Abdominal Stent Graft System

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Herz- und Gefässzentrum Bad Bevensen, Bad Bevensen, Germany

Loading trial locations.

About this study

A prospective, consecutively enrolling, non-randomized multi center clinical evaluation of the safety and performance of the TriVascular AAA Stent Graft when used in the treatment of patients with Abdominal Aortic Aneurysms (AAA).

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patient is > 18 years of age
  • Patients who are male or non-pregnant female (females of child bearing potential must have a negative pregnancy test prior to enrollment into the study)
  • Patient has signed an Ethics Committee (EC) approved Informed Consent Form
  • Patient is considered by the treating physician to be a candidate for elective open surgical repair of the AAA (i.e., category I, II, or III per American Society of Anesthesiology (ASA) classification; refer to Appendix III: ASA Classification System). ASA category IV patients may be enrolled provided their life expectancy is greater than 1 year.
  • Patient has an infrarenal abdominal aortic aneurysm that meets at least one of the following:
  • Abdominal aortic aneurysm ≥5.0 cm in diameter
  • Aneurysm has increased in size by 0.5 cm in last 6 months.
  • Maximum diameter of aneurysm exceeds 1.5 times the transverse dimension of an adjacent non-aneurysmal aortic segment
  • Patient has patent iliac or femoral arteries that allow endovascular access with the TriVascular AAA device.
  • Patient has a suitable non-aneurysmal proximal aortic neck length of ≥ 7 mm inferior to the most distal renal artery ostium.
  • Patient has a suitable non-aneurysmal distal iliac artery length (seal zone) of ≥10 mm. The resultant repair should preserve patency in at least one hypogastric artery.
  • Patient has a suitable non-aneurysmal proximal aortic neck luminal diameter between 16 and 30 mm.
  • Patient has suitable non-aneurysmal distal iliac luminal diameters between 8 and 20 mm.
  • Patient meets the following anatomic criteria: the distance from the most distal renal artery to most superior internal iliac artery measurement is at least 13 cm.
  • Patient has juxtarenal aortic neck angulation ≤ 60º if proximal neck is ≥10 mm and ≤ 45º if proximal neck is <10 mm.
  • Patient must be willing to comply with all required follow-up exams.

Exclusion criteria

  • Patient has a dissecting aneurysm
  • Patient has an acutely ruptured aneurysm
  • Patient has an acute vascular injury
  • Patient has a need for emergent surgery
  • Patient has a known thoracic aortic aneurysm or dissection.
  • Patient has a mycotic aneurysm or has an active systemic infection
  • Patient has unstable angina (defined as angina with a progressive increase in symptoms, new onset at rest or nocturnal angina, or onset of prolonged angina)
  • Patient has had a myocardial infarction (MI) and/or stroke (CVA) within the past 6 months.
  • Patient has a major surgical or interventional procedure planned ≤30 days of the AAA repair.
  • Patient has history of connective tissue disease (e.g., Marfan's or Ehler's-Danlos syndrome).
  • Patient has history of bleeding disorders or refuses blood transfusions.
  • Patient has dialysis dependent renal failure or baseline serum creatinine level >2.0 mg/dl
  • Patient has a known hypersensitivity or contraindication to anticoagulation or contrast media that is not amenable to pre-treatment.
  • Patient has a known allergy or intolerance to polytetrafluoroethylene (PTFE), PEG-based polymers or nitinol.
  • Patient has a body habitus that would inhibit X-ray visualization of the aorta
  • Patient has a limited life expectancy of less than 1 year
  • Patient is currently participating in another investigational device or drug clinical trial
  • Patient has other medical, social or psychological conditions that, in the opinion of the investigator, preclude them from receiving the pre-treatment, required treatment, and post-treatment procedures and evaluations.

Treatment and study plan

Implant of Ovation Abdominal Stent Graft System

Device

Single occurrence permanent implant of AAA device.

Primary outcomes

  1. The primary safety endpoint is defined as the proportion of subjects who experience a Serious Adverse Event within 30 days of the initial procedure.

    Time frame: 30-Days

    The safety of the TriVascular AAA Stent Graft will be determined by comparing the rate of Serious Adverse Events in the Treatment Group against the rate of Serious Adverse Events in the Control Group within 30 days of the initial procedure.

Secondary outcomes

  1. To evaluate the performance of the TriVascular AAA Stent-Graft System within 30 days of the initial procedure.

    Time frame: 30-Days

    The performance of the TriVascular AAA Stent-Graft System will be determined by comparing the composite success rate in the Treatment Group against an estimated success rate of 80% in the Control Group within 30 days of the initial procedure. The composite success rate is defined as:

    • Technical Success, defined as successful delivery and deployment of one aortic body and two iliac limbs.
    • Absence of Type I and III endoleaks
    • Absence of Rupture
    • Absence of Conversion to Open Surgical Repair
    • Freedom from mortality

Sponsors and collaborators

Lead sponsor

TriVascular, Inc.

Industry

Registry information

Official study title

A Clinical Study to Evaluate the Safety and Performance of the TriVascular Ovation™ Abdominal Stent Graft System

Important dates

Study start
2010
Primary completion
2010
Study completion
2015
First posted
Apr 2, 2010
Registry last updated
Jun 4, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.