PresVIEW Scleral Implants
DeviceSubjects are implanted with the PresView Scleral Spacing Implants and followed for 24 months.
NCT Number: NCT01166568
The primary objective of this study is to evaluate the safety and effectiveness of the PresVIEW™ Scleral Implant (PSI) for the improvement of near visual acuity in presbyopic patients.
Looking for future studies?
Notify Me50 year–60 year
All sexes
Interventional
Not applicable
Boxer-Wachler Vision Institute, Beverly Hills, California, United States
The clinical investigation of the PresView™ Scleral Implant (PSI) is a prospective multicenter clinical trial in which a total of 330 subjects were enrolled and implanted with the PSI and followed for a period of 24 months at up to 14 clinical sites.
An additional randomized sub-study was performed on 48 subjects with the PSI Second Generation PresView™(SGP) implant, model number SGP-046. A 2:1 randomization was used to determine the 32 subjects assigned to be implanted and the 16 subjects assigned to the observation/deferred implantation control arm. Subjects randomized to the deferred implantation control arm were eligible to receive the PSI after completion of the 6 months of observation in the study.
All subjects interested in participating in the study were screened for eligibility, and informed consent was obtained from those who meet the inclusion/exclusion criteria. Eligible subjects were examined preoperatively to obtain a medical history and baseline ocular data.
Effectiveness: The following tests/measurements used to evaluate the effectiveness of the PSI procedure will be included in the study protocol:
The PSI procedure will be defined as successful if a logMar equivalent to Snellen 20/40 or better (logMar 0.3) is achieved in ≥ 75% of patients or if 75% of patients realize an improvement ≥ 2 lines in distance corrected near visual acuity (DCNVA). For most patients, this will translate into the ability to read the majority of newspaper and magazine print without a near optical aid.
Safety: Primary safety outcomes for safety will include:
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Subjects are implanted with the PresView Scleral Spacing Implants and followed for 24 months.
Time frame: From date of baseline measurement until the date of study withdrawal or study completion, whichever came first, assessed up to 2 years.
Measurement of the Distance Corrected Near Visual Acuity at 40 centimeters achieving 20/40 or better or improvement of 2 or more lines at 24 months for the primary eye.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
The point estimate were calculated for clinically significant adverse events (SAE's) at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis. Point estimates were calculated for the incidence of Anterior Segment Ischemia at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2% NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis. Point estimates were calculated for the incidence of Decrease in BCDVA > 2 lines at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis. Point estimates were calculated for the incidence of Decrease in BCNVA > 2 lines at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis (i.e. both primary and fellow eyes). Point estimates were calculated for the incidence of (IOP Increase > 10 mmHg) or (IOP > 25 mmHg) at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis. Point estimates were calculated for the incidence of Chronic Inflammation at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Time frame: From initiation of the implantation procedure (operative day) until study completion at 24 months or withdrawal from the study.
All primary and fellow eyes implanted with the PSI were included in the safety analysis. Point estimates were calculated for the incidence of (Axial Length Increase >= 0.20) AND (Myopic Shift > 0.50D MRSE) at the end of the study. Incidence of adverse event totals should not to exceed 5%, and the incidence of each event should not exceed 1-2%.
NOTE: Because the secondary outcome is a safety measure, both primary and fellow eyes were used for analysis.
Refocus Group, Inc.
Industry
A Prospective, Multicenter Clinical Trial Of the PresVIEW Scleral Implant (PSI) For the Improvement Of Near Visual Acuity In Presbyopic Patients
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05656027
Eye Diseases, Myopia
Chandler, Arizona, United States
View Trial DetailsNCT05753189
Eye Diseases, Myopia
Dothan, Alabama, United States
View Trial DetailsNCT06869044
Eye Diseases, Hyperopia
Berkeley, California, United States
View Trial DetailsNCT05728944
Eye Diseases, Myopia
Dothan, Alabama, United States
View Trial Details