Federal University Of São Paulo - Gastroenterology
São Paulo, Brazil
NCT Number: NCT00725244
Aim:
To compare the efficacy, safety and number of sessions of bipolar eletrocoagulation (BEC) and argon plasma coagulation (APC) in the management of the bleeding telangiectasias from chronic radiation coloproctopathy (CRCP).
Methods:
Thirty patients with active bleeding from telangiectasias were enrolled in two groups (15 BEC and 15 APC) and classified according of Saunders score. BEC settings were 50 W and APC settings were 40 W and 1.0 l/min. Colonoscopy was the first exam to rule out synchronous lesions and follow-up was performed with sigmoidoscopies. Clinical cure was defined as cessation of bleeding and endoscopic cure was determined by absence of telangiectasias. Failure was defined whenever more than 7 sessions or other therapy was necessary.
Looking for future studies?
Notify Me18 year–88 year
All sexes
Interventional
Phase 4
São Paulo, Brazil
Thirty patients with active and chronic hematochezia from radiation telangiectasias were randomized in two groups :
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Bipolar Eletrocoagulation was performed with a high-frequency electrosurgical generator (ERBE® ICC 200 Eletromedizin, Tubingen, Germany), using Gold probe (Wilson- Cook®) with 7 Fr diameter and 300 cm length. The power setting was 50 W. Coagulation of each telangiectasia was achieved with the probes by applying light pressure directly on the telangiectasia.
Other names: - bipolar probe, - bipolar electrocoagulation
Argon plasma coagulation was delivered using a "spray-painting" technique, with short applications at 40 W power with a gas flow of 1.0l per minute. APC equipment was an argon delivery unit (ERBE® ICC 300) coupled a high frequency surgery unit (ERBE® ICC 200). Only the end-firing probe with 2.3 mm and 220 cm length was used. The probe was purged with argon, tested and passed though the endoscope until it extends approximately 1 cm from the tip. The probe was hold just above the mucosal surface and the contact was avoided. During the procedure periodic suction was made to prevent over-distention with gas and consequently patient discomfort.
Other names: Argon Laser, Argonio
Time frame: 6 months
Time frame: 6 months
Federal University of São Paulo
Other
A Randomized Prospective Study of Endoscopy Bipolar Eletrocoagulation and Argon Plasma Coagulation of Chronic Rectal Bleeding From Radiation Telangiectasias
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02104271
Cardiovascular Diseases, Digestive System Diseases
Barretos, São Paulo, Brazil
View Trial DetailsNCT02972736
Radiation Injuries, Wounds and Injuries
View Trial DetailsNCT04059809
Breast Cancer, Breast Diseases
Itajaí, Santa Catarina, Brazil
View Trial DetailsNCT03667859
Radiation Injuries, Wounds and Injuries
Haifa, Israel
View Trial Details