Connex CVSM with Timmy3 module
DeviceConnex Vital Signs Monitor with the Timmy3 Investigational module
NCT Number: NCT06056011
The purpose of this document is to validate the clinical accuracy of the TIMMY3 thermometry module, integrated into host device CVSM, according to ISO 80601-2-56:2017 + A1 2018.
Looking for future studies?
Notify Me1 day–99 year
All sexes
Interventional
Not applicable
First Georgia Physicians Group, Fayetteville, Georgia, United States
The purpose of this document is to validate the clinical accuracy of the TIMMY3 thermometry module, integrated into host device CVSM, according to 80601-2-56:2017 + A1 2018.
This study is designed to provide supporting documentation for the TIMMY3 module operating with SureTemp Plus algorithms. Testing will be performed with a production equivalent CVSM modified to use TIMMY3 (TIMMY3 investigational device), running the SureTemp Plus algorithms. The TIMMY3 module is intended to be integrated into additional Welch Allyn devices such as, but not limited to, the Connex Spot Monitor and Spot 4400. The process described below will be followed in accordance with the method called out in ISO 80601-2-56:2017 + A1:2018.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Note: The study subject will be excluded if the reference temperature does not stabilize after the 3 -5 minute monitoring time.
Connex Vital Signs Monitor with the Timmy3 Investigational module
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Clinical bias is the mean difference between the first device under test (DUT) output temperature (T1) and the corresponding reference clinical thermometer (TRCT) within the ISO-compliant analysis set and age group, calculated per ISO 80601-2-56:2017 + A1:2018 (refer to ISO formula [2]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Limits of Agreement represent the magnitude of potential disagreement between outputs of two thermometers and are defined as 2 × the standard deviation (SD) of the temperature differences (T1 - TRCT) within the ISO-compliant analysis set and age group, as specified in ISO 80601-2-56:2017 + A1:2018 (refer to ISO formulas [3] and [4]).
Time frame: 20-30 minutes
Clinical repeatability is defined per ISO 80601-2-56:2017 + A1:2018 as the pooled within-subject standard deviation (SD) of triplicate device-under-test output temperatures for each anatomical site (refer to ISO formulas [5] and [6]). This outcome reports that SD as a single numeric value (°C); because it is a dispersion statistic (not a mean), no central-tendency measure applies. Higher SD indicates lower repeatability.
Baxter Healthcare Corporation
Industry
TIMMY3 80601-2-56:2017 + A1 2018 Clinical Accuracy Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04925401
Body Temperature Changes, Febrile Illness
Paris, France
View Trial DetailsNCT05045547
Febrile Illness
Battambang, Cambodia
View Trial DetailsNCT04760678
Febrile Illness
Kimpese, Bas-Congo Province, Democratic Republic of the Congo
View Trial DetailsNCT05850013
Cardiovascular Diseases, Febrile Illness
Sunderland, United Kingdom
View Trial Details