Capture rate following void (captured as % of urine captured by device and collected in canister, measured by weight).
Prior to each nightly use, the weight of the empty urine collection canister and the weight of a dry absorbent pad are collected. Participants use the device overnight while sleeping with the absorbent pad underneath them. The next morning, the post-void weight of the urine collection canister and the post-void weight of the absorbent pad are collected.
Capture rate is calculated as the weight of captured urine (post-void canister weight - pre-void canister weight) / total urine volume (captured urine (post-void canister weight - pre-void canister weight) + leaked urine (post-void pad weight - pre-void pad weight)) * 100.