Hospital Vall d'Hebron
Barcelona, 08035, Spain
NCT Number: NCT04380493
The purpose of this study is to assess the reliability, reproducibility and accuracy of the paediatric probe of transient elastography in detecting liver fibrosis in children, besides its limitations and side effects. At the same time, to assess whether indirect fibrosis markers are a valid tool to detect absence or mild fibrosis in paediatric patients
Looking for future studies?
Notify MeUp to 18 year
All sexes
Observational
Barcelona, 08035, Spain
A retrospective cohort study, including all patients younger than 18 years of age with chronic liver disease who underwent a transient elastography, was performed. The study was carried out in the joint Unit of Paediatric Complex Hepatology and Liver Transplant, encompassing two third level paediatric hospitals, between 2015 and 2019. The study protocol was approved by local ethic committee (identifier code: PR(AMI)147/2019) and informed consent was signed by patient's legal guardians and patients over 16 years old.
Liver stiffness was assessed with FibroScan® (Echosens, France, model 502, class IIa, year 2010), which was performed by an experienced operator to avoid interobserver bias and in the same standardized conditions (fasting, decubitus position, right arm in maximal abduction, transducer placed over the right lobe). S probe (model 8; 5 MHz; diameter 5 mm), that measures depths ranging from 15-50 mm (S1: 15-40 mm; S2: 20-50 mm), was selected if chest circumference (CC) < 75 cm (S1: ≤ 45 cm; S2: 45-75 cm). M probe (3.5 MHz; diameter 7 mm), measuring depths range 35-75mm, was used if CC ≥ 75 cm. Both probes met the calibration terms. Elastographic parameters studied were: stiffness (KPa), interquartile range (IQR) and success rate (SR). Adult-validated liver stiffness measurement ranges to classify the fibrosis degree were taken as a reference: F0-F1 (≤7.6 KPa), F2 (7.7-9.4 KPa), F3 (9.5-14 KPa) and F4 (>14 KPa). The examination was successful when the median of at least 10 valid values, had a SR higher than 60% and with an IQR/LSM lower than 30%.
The following sociodemographic, clinical and analytical data were collected: age, sex, underlying liver disease and analytical parameters (alanine aminotransferase (ALT) U/L, aspartate aminotransferase (AST) U/L, gamma glutamyl transferase (GGT) U/L, total bilirubin (mg/dl), cholesterol (mg/dl), platelets (109/L), prothrombin time (%)). The following liver fibrosis test Scores were calculated using the obtained data:
Eighteen clinically indicated liver biopsies were performed and staged according METAVIR score (F0-F4). Blood test, fibrosis scores and biopsies were performed within 6 months of transient elastography. No clinical or analytical changes were observed during this period.
The Faces Pain Scale was used for assessing pain severity. Statistical analysis: Kolmogorov-Smirnov test checked out if variables followed a normal distribution. Continuous variables are presented as mean ± standard deviation or median (interquartile range) and categorical variables as number of subjects and percentage. Correlation between continuous variables were calculated using Pearson's and Spearman's correlation coefficient. The variance's analysis (ANOVA) and the Kruskal Wallis test were used to compare qualitative with quantitative variables. Two-tailed p-Values of <0.05 were considered as statistically significant. Sensitivity, specificity, positive predictive value and negative predictive value were calculated in the assessment of FIB-4 and APRI as a diagnostic test to detect absence or mild fibrosis (F0-F1), using the results obtained by TE as a reference. The statistical analyses were performed with SPSS software (v. 21, SPSS Inc., Chicago, IL, USA) and Prism (v. 7.04 GraphPad Software Inc.).
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Other names: Compare results with liver biopsy, Calculate indirect fibrosis markers (APRI and FIB-4)
Time frame: 4 years
To describe our experience with FibroScan's® pediatric (S) probe; its reliability, reproducibility, and accuracy in detecting fibrosis, as compared to M probe and histology, as well as its limitations and side effects
Time frame: 4 years
To assess whether indirect fibrosis markers (APRI and FIB-4) are a valid tool to detect absence or mild fibrosis in paediatric patients
Hospital Vall d'Hebron
Other
Special Considerations for the Use of the Transient Elastography Paediatric Probe, Compared to the Standard M Probe, Indirect Serological Markers and Histology
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02372682
Cirrhosis, Digestive System Diseases
St Louis, Missouri, United States
View Trial DetailsNCT06463223
Attenuation Imaging, Body Weight
Gothenburg, Västra Götaland County, Sweden
View Trial DetailsNCT02446665
Bile Duct Diseases, Biliary Tract Diseases
Toronto, Ontario, Canada
View Trial DetailsNCT07127302
Arm Injuries, Elastography
Karaman, Merkez, Turkey (Türkiye)
View Trial Details