NCT Number: NCT01913561

The Use of Master Caution Garment for Continuous Monitoring for Arrhythmia Detection

The Master Caution Garment is embedded with 10/13 textile dry electrodes that enable 12/15-lead ECG.

The electrodes are dry, textile ECG electrodes. The garment is designed to position the electrodes in the appropriate anatomical locations, without any special guidance. The dry electrodes are made from unique yarns with silver embedded into the garment using a proprietary technique developed by HealthWatch (HW)that obviates the need for skin preparation or the application of fluids. The garment can be connected to any standard ECG device available in the hospitals.

The purpose of this study is to evaluate the safety and effectiveness of the Master Caution Garment for continuous monitoring of ECG signal compared to standard gel electrodes.

The following endpoints will be evaluated in subjects participating in the study:

Primary:

Compare the quality of the ECG signal of HW textile electrodes compared to gel electrodes using standard ECG devices.

Secondary:

Compare the quality of the ECG signal of HW textile electrodes compared to gel electrodes when event is detected by the ECG devices.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–90 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Rambam medical center

Haifa, 9602, Israel

About this study

Part A - ECG in Rest

  • In the beginning of the study, the following parameters will be measured and documented in the CRF: blood pressure, body temperature, respiratory rate and oxygen saturation in the blood.
  • The patient will be dressed with the Master Caution Garment and connected to a 12 lead ECG device. The procedure takes 5 minutes including data printing.
  • A twelve lead ECG with gel electrodes will be done and data printed. The procedure takes 5 minutes. The patient will stay connected to the monitor for the following hour.
  • After an hour the data from the gel electrodes will be printed again.
  • The patient will be dressed again with the Master Caution Garment and connected to a 12 lead ECG device. The procedure takes 5 minutes including data printing.
  • The following parameters will be compared between the gel ECG and the Garment ECG:
  • Heart Rate (HR)
  • P wave width and amplitude
  • QRS complex width and amplitude
  • T wave width and amplitude
  • RR interval
  • PR interval
  • QT interval

Part B - ECG in motion

  • In the beginning of the study, the following parameters will be measured : blood pressure, body temperature, respiratory rate and oxygen saturation in the blood.
  • Four (4) gel electrodes will be placed on the patient on the following places - Left Arm (LA), Right Arm (RA), Left Leg (LL) and Right Leg (RL).
  • The patient will be dressed with the Master Caution Garment (MCG)on the 4 gel electrodes that were already placed.
  • The same 4 electrodes in the garment LA, RA, LL and RL will be connected to a standard of care ECG device.
  • The 4 gel electrodes will be connected to a standard of care ECG device.
  • The ECG devices will be recording the ECG signal from the two sets of electrodes for 24 hours.

Every hour in the first 6 hours the data will be printed from both devices. In every event detected the data will be printed from both devices. the data will be evaluated for quality comparison between the gel and the textile electrodes using the hospital ECG device.

  • In first 6 hours for every hour and for each event detected by either the gel electrodes or the MCG textile electrodes the following parameters shell be evaluated at the end of the study from the hospital ECG printouts for the Master Caution Garment and for the gel electrodes:
  • Heart Rate (HR)
  • P wave width and amplitude
  • QRS complex width and amplitude
  • T wave width and amplitude
  • RR interval
  • PR interval
  • QT interval After first 6 hours for each event detected by either the gel electrodes or the MCG textile electrodes the above mentioned parameters shell be evaluated at the end of the study from the hospital ECG printouts for the Master Caution Garment and for the gel electrodes.

The goal of the current study is to evaluate the safety of the MCG and compare ECG signals from the Master Caution Garment with the ECG signals from gel electrodes.

This is a prospective, comparative study.

For each time interval and each event detected by the ECG devices, the analysis will determine its category as follows:

  • Blinded cardiologist's assessment whether at that time interval the patient suffered from a disorder. This assessment will be based on hospital's recorded data with the gel electrodes.
  • The Master Caution Garment result for that time interval will be categorized as follows:
  • True Positive (TP) = MCG Alert was right (based on GS)
  • False Positive (FP) = MCG Alert was wrong
  • True Negative (TN) = No MCG Alert when indeed alert was not necessary
  • False Negative (FN) = No MCG Alert but there was a disorder case

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age>18 years.
  • Suspected for arrhythmia or have evidence of arrhythmia disorder.
  • Needs continuous, in-hospital ECG monitoring.
  • Ability to give informed consent.

Exclusion criteria

  • Patients with known allergy to silver.
  • Patients with recent trauma to chest wall.
  • Pregnant or lactating woman.
  • Patients with chest deformity precluding placement of the garment.
  • Patients participating in another ongoing trial.

Treatment and study plan

Master Caution Garment

Device

Master Caution Garment is embedded with 10/13 textile dry electrodes that enable 12/15-lead ECG.

The electrodes are dry, textile ECG electrodes.

Other names: MCG

ECG gel electrodes

Device

Standard gel electrodes that are been in use in hospital for ECG monitoring.

Primary outcomes

  1. From the ECG record collect the Heart Rate HR.

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  2. From the ECG record collect the P Wave width (sec.) and the amplitude (mV).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  3. From the ECG record collect the QRS Complex width (sec.) and the amplitude (mV).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  4. From the ECG record collect the T Wave width (sec.) and the amplitude (mV).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  5. From the ECG record collect the RR Interval width (sec.).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  6. From the ECG record collect the PR Interval width (sec.).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

  7. From the ECG record collect the QT Interval width (sec.).

    Time frame: 0,1,2,3,4,5,6 hours and in each arrythemia event detected durring 24 hours from the start of the ECG recording.

Sponsors and collaborators

Lead sponsor

HealthWatch Ltd.

Industry

Registry information

Official study title

The Use of Master Caution Garment for Continuous Monitoring for Arrhythmia Detection Compared To Standard Gel-Electrode-Based ECG

Important dates

Study start
2013
Primary completion
2017
Study completion
2017
First posted
Aug 1, 2013
Registry last updated