SAVVY guidewire
DeviceThe SAVVY guidewire will be used for rapid pacing runs, valve delivery and for left ventricular pressure measurements pre- and post-THV deployment
NCT Number: NCT05082337
TAVR patients exhibiting post-procedural residual AR had higher mortality and hospitalization rates due to heart failure, with the extent of this association increasing proportionally to the severity of the regurgitation. Optimizing transcatheter valve performance by intra-procedural hemodynamic evaluation of AR and residual transvalvular gradient remains of high clinical importance.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
ICM, Montreal, Quebec, Canada
Continuous technological development and procedural refinements are important to further reduce TAVR peri-procedural complications, facilitate the minimalist strategy and improve clinical outcomes following TAVR. Residual aortic regurgitation (AR) has been one of the main drawbacks of TAVR. TAVR patients exhibiting post-procedural residual AR had higher mortality and hospitalization rates due to heart failure, with the extent of this association increasing proportionally to the severity of the regurgitation.
Various actions could be undertaken to minimize AR and its consequences, either intraprocedural (balloon post-dilation, implantation of a second valve) or during follow-up (diuretic treatment, closest clinical follow-up or percutaneous leakage closure). Thus, an accurate diagnosis of the presence and severity of residual AR post-TAVR is key to implement the proper measures and optimize clinical outcomes.
The use of cardiac imaging remains the gold standard for evaluating AR post-TAVR, limitations of contrast amount along with the subjectivity of AR evaluation by aortic angiography, and the challenges of echocardiography at the time of the TAVR procedure represent a significant drawback. Thus, alternative tools for evaluating the presence and severity of AR would be very helpful in this setting.
Current data supports the implementation of hemodynamic measurements during TAVR procedures in order to improve the clinical decision-making process following valve implantation. However, these actions entail an inherent risk associated with the exchange of catheter-wires. The use of a support guidewire with pressure measurement capabilities would facilitate the hemodynamic evaluation of transcatheter valve performance in a safer and more rapid manner.
The new SAVVY guidewire, with both dedicated pacing properties and allowing a continuous hemodynamic pressure monitoring during the procedure, is a unique system in the field and may represent an important step forward in the process of optimizing the TAVR procedure while facilitating procedural steps.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The SAVVY guidewire will be used for rapid pacing runs, valve delivery and for left ventricular pressure measurements pre- and post-THV deployment
Time frame: Periprocedural
Presence of major complications related to the SAVVY guidewire including (i) guidewire kink hindering or preventing the advancement of the transcatheter valve system
Time frame: Periprocedural
Major vascular complications related to the SAVVY guidewire are defined as stroke, bleeding or left ventricular perforation
Time frame: Periprocedural
Pacing capture failure translating into valve malpositioning. Valve malpositiong will be evaluated by a landing site to low or too high, leading to hemodynamically unfavorable results.
Time frame: Periprocedural
Pacing capture failure translating into valve embolization
Time frame: Periprocedural
Effective rapid pacing will be defined as an adequate ventricular pacing capture for a minimum of 10 seconds, with no capture loss, and leading to a reduction of aortic pressure of at least 50%, with/or a systolic pressure value <60 mmHg. Efficacy will be assessed by the physician.
Time frame: Periprocedural
Accurate ventricular pressure will be defines as a pressure wire measurements similar (differences <5 mmHg) to those obtained simultaneously with a pigtail catheter in the same cavity or vascular segment (differences <5 mmHg in SBP, differences <5 mmHg in LVEDP).
Institut universitaire de cardiologie et de pneumologie de Québec, University Laval
Other
The SAVVY Guidewire in Transcatheter Aortic Valve Replacement Procedures: A Feasibility Study
Acronym: SAVVY
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06896227
Aortic Valve Disease, Aortic Valve Stenosis
Düsseldorf, Germany
View Trial DetailsNCT04437303
Aortic Valve Disease, Aortic Valve Stenosis
Aalst, Belgium
View Trial DetailsNCT03807921
Aortic Valve Disease, Aortic Valve Insufficiency
Montevideo, Uruguay
View Trial DetailsNCT04557345
Aortic Valve Calcification, Aortic Valve Disease
Mexico City, Mexico
View Trial Details