NCT Number: NCT07576946
This study aims to evaluate the functional performance and feasibility of the Saturista system, a novel device designed for automated oxygen flow regulation based on continuous monitoring of oxygen saturation (SpO₂). The system integrates real-time physiological monitoring, wireless data transmission, and centralized visualization to support timely detection of oxygen desaturation and clinical decision-making.
A pilot feasibility study will be conducted in healthy volunteers using a scenario-based protocol, including baseline monitoring, controlled desaturation, probe disconnection, and signal recovery. The primary objective is to assess the system's ability to generate alarms in response to changes in SpO₂ and to display data in real time. Secondary objectives include evaluation of response time, signal management, and overall system stability.
This study is active but is not currently recruiting participants.
Notify Me18 year and older
All sexes
Observational
Istanbul, Turkey (Türkiye)
The Saturista system is a novel, integrated platform designed to support automated oxygen therapy management through continuous monitoring of oxygen saturation (SpO₂) and real-time decision support. The system combines pulse oximetry-based sensing, wireless data transmission, a cloud-based data infrastructure, and a web-based user interface within a distributed closed-loop control architecture.
In conventional clinical practice, oxygen therapy is typically adjusted manually based on intermittent assessment of oxygen saturation. This approach may delay recognition of physiological deterioration and timely intervention. The Saturista system addresses this limitation by enabling continuous monitoring, automated alarm generation, and centralized visualization of patient data at the nursing station.
The system consists of three main components: (i) a pulse oximetry-based monitoring module, (ii) an embedded device with wireless communication capabilities that regulates oxygen flow via a pneumatic valve mechanism, and (iii) a cloud-based software platform that processes physiological data and provides real-time visualization and alarm management.
This study is designed as a pilot feasibility study conducted in healthy volunteers to evaluate the technical and functional performance of the system. A scenario-based testing protocol will be applied, including baseline monitoring, controlled desaturation through short-duration breathing maneuvers, probe disconnection, and signal recovery. These scenarios simulate common clinical monitoring conditions and allow assessment of system responsiveness and reliability.
The primary outcome measures include the system's ability to detect decreases in SpO₂ and generate appropriate alarms, as well as its capacity to display physiological data in real time. Secondary outcomes include system response time, detection of signal loss, recovery performance, and overall operational stability.
The findings of this study are expected to provide preliminary evidence regarding the feasibility and functional performance of the Saturista system and to inform the design of future clinical studies in patient populations.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of measurements successfully displayed on the user interface in real time during all testing scenarios.
Time frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between a decrease in SpO₂ below the predefined threshold and activation of the system alarm.
Time frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Time (in seconds) between SpO₂ measurement and visualization on the user interface.
Time frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Proportion of scenarios in which the system successfully resumes valid data acquisition and terminates alarms after probe reconnection.
Time frame: Day 1, during multiple predefined simulation scenarios applied sequentially within a single testing session
Frequency of technical issues or system failures observed during testing sessions.
Koç University
Other
Development and Pilot Evaluation of an Artificial Intelligence-Assisted System for Automated Oxygen Flow Regulation: The Saturista Device
Acronym: SATURİSTA
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06037915
Child, Laryngoscopy
Edegem, Antwerp, Belgium
View Trial DetailsNCT06816706
Critical Care, Mechanical Ventilation
Beijing, Beijing Municipality, China
View Trial DetailsNCT03431818
Oxygen Therapy
Miami, Florida, United States
View Trial DetailsNCT06816745
Critical Care, Oxygen Therapy
Beijing, Beijing Municipality, China
View Trial Details