Bangkok Metroplitan Administration and Vajira Hospital
Bangkok, 10300, Thailand
NCT Number: NCT04454606
Filter facepiece respirator (FFR) is a critical equipment to prevent the transmission of respiratory tract infection disease especially the dreadful corona virus 2(SARs-CoV-2).The N95 mask is the prototype of high efficiency protective device and can effectively protect airborne pathogens of less than 0.3 μm by more than 95%. It is tightly fit and had high filtration capacity. The widespread pandemic of COVID-19 leads to greater requirement of FFR. A rising in demand would greatly exceed current productive capabilities and stockpiles and would almost certainly result in a robust shortage.
In order to solve these problems, our team had invented a new type of half-piece respirator made from silicone and assembled with hepa or elastostatic filter . A variety of methods have been used to evaluate this new device, including qualitative fit test with Bitrex® test kit and filtration test.
Looking for future studies?
Notify Me18 year–60 year
All sexes
Interventional
Not applicable
Bangkok, 10300, Thailand
As required by Occupational Safety and Health Administration OSHA)[7], the halfpiece respirators need to pass the fittestto identify those individuals who do not achieve the good fit necessary for adequate protection. The performance of this respirator was also determined by measuring percent leakage under constant airflow[8]. The purposes of this study was to evaluate the fitting characteristicof our newly invented silicone N99 respirator and the performance against 30 µm particles using NaCL aerosols. The study was conducted to achieve two specific research objectives
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Bitrex Fit Test The Bitrex test uses a person's ability to test a bitter solution to determine whether a respirator fits properly[11]. Each subject was given a taste-threshold screening test prior to each fit test to ensure that he or she could taste. This process was done without subject wearing a respirator. After passing the sensitivity test, the subject will process to 7 steps of fit test as following
Respirator Performance
The real time respirator performance test method was developed using MT-05U machine (SIBATA model, Saitama, Japan), which measure particle concentrations of diameter 0.03 - 0.06 µm using and particle generator laser beam scattering particle counter that measure particles outside and inside the mask. The ratio between these two values reflected % leak as following:
Study protocol
Time frame: Day 1
pass or fail
Time frame: Day 1
percent leakage less than1 percent
Bangkok Metropolitan Administration Medical College and Vajira Hospital
Other Gov
The New Silicone N99 Half-Piece Respirator:VJR-NMU N99:New and Effective Tool to Prevent COVID-19
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06821139
COVID, COVID - 19
Surrey, British Columbia, Canada
View Trial DetailsNCT04407143
COVID, COVID-19
A Coruña, Spain
View Trial DetailsNCT04394390
Avitaminosis, COVID
Bursa, Dogankoy, Turkey (Türkiye)
View Trial DetailsNCT04395599
COVID, COVID-19
Lille, France
View Trial Details