Skip to main content
OpenTrials
Completed

NCT Number: NCT00766935

The Equivalency of the Imp SFB7+ (L-Dex U400) With the Imp XCA Device

The purpose of this project is to prove that the Imp™SFB7+ (L-Dex U400) is equivalent to the Imp™ XCA device in assessing Lymphoedema. The Imp™ XCA device is a simple device capable of assessing Lymphoedema of a single arm. The Imp™SFB7+ (L-Dex U400) is a more complex device capable of assessing Lymphoedema in a larger population of Lymphoedema sufferers. The Imp™SFB7+ (L-Dex U400) will be developed to assess patients with Lymphoedema in their arms or their legs. The Imp™ XCA and Imp™ SFB7 are both TGA approved devices. The Imp™ SFB7+ (L-Dex U400) is similar to the Imp™ SFB7 the changes are to the software and user interface.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–75 year

Sex eligibility

Female

Study type

Observational

Primary location

Queensland Lymphoedema and Breast Oncology Physiotherapy

Brisbane, Queensland, 4032, Australia

About this study

When Lymphoedema is present, lymph and other fluids build up in the interstitial spaces of the tissues. This results in an overall increase in the total amount of extracellular fluid (ECF) in the limb, causing swelling. This can be documented by measuring the impedance (opposition) to a low frequency current that has been passed into the limb. Low frequency current travels predominantly through the ECF, where the Lymphoedema manifests. As the fluid builds up in the limb, the impedance to the current decreases and it is in this way that low frequency bioimpedance is able to assess Lymphoedema.

Multi-frequency bio-impedance analysis otherwise known as Bioimpedance Spectroscopy (BIS) has been reported to be effective for the measurement of ECF and sub clinical changes in ECF to predict the onset of Lymphoedema in the arms in studies conducted by Cornish et al. It has been reported by Warren et al that BIS can be used as a reliable and accurate tool for documenting presence of lymphoedema in patients with wither upper- or lower-extremity swelling.

The Imp™ XCA device uses an "impedance ratio" methodology to assess unilateral Lymphoedema of the arm. By this method the unaffected arm acts as an internal and subject specific control. The Imp SFB7+ (L-Dex U400) uses the same principles but at a lower frequency. This study is designed to show the equivalence of the 2 devices in assessing unilateral Lymphoedema of the arm.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

Groups 1 and 2

  • Be female between the ages of 18-75 years.
  • Self-describe general health as satisfactory.
  • Understand the proposed study and be willing and fully able to comply with the study procedures.
  • Be a willing participant and be capable of giving and has given informed written consent for entry into the study.

Group 1

  • Have been previously diagnosed, by current clinical practice, as having unilateral Lymphoedema of the arm of any severity.
  • Have had a mastectomy or breast conservation surgery with axillary sampling or dissection, with or without adjuvant therapy.

Exclusion criteria

  • Have a known heart condition or an implantable device such as a pacemaker or ICD.
  • Have a metallic surgical implant (e.g. total hip replacement) not including small implants such as sternal wires or surgical staples.
  • Suffer from a renal disorder.
  • Be taking diuretic medications.
  • Have consumed large amounts of alcohol or caffeinated beverages within six hours of the study.
  • Have undertaken excessive exercise within two hours of BIA.
  • Have a reported fever of > 38oC at time of screening.
  • Be currently in the fourth week of the menstrual cycle.
  • Be pregnant or currently breastfeeding.
  • Be a relative of any member of study staff or is an employee or a relative of an employee of ImpediMed Ltd.

Treatment and study plan

Primary outcomes

  1. Substantial equivalence of the SFB7+ (L-Dex U400) to XCA in the assessment Lymphoedema

    Time frame: Single point measure

    Observation of difference between L-Dex values of lymphoedema population compared to healthy individuals

Sponsors and collaborators

Lead sponsor

ImpediMed Limited

Industry

Registry information

Important dates

Study start
2007
Primary completion
2007
Study completion
2008
First posted
Oct 6, 2008
Registry last updated
Aug 27, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.