Ling Liu
Nanjing, Jiangsu, 210009, China
NCT Number: NCT05989971
In the design of extracorporeal carbon dioxide removal (ECCO2R) combined with continuous renal replacement therapy (CRRT) equipment, in model of continuous veno-venous hemofiltration (CVVH) , the HCO3- concentration in the pre membrane lung blood is diluted by the replacement solution, and a decrease in HCO3- leads to a decrease in PCO2. On the other hand, in continuous veno-venous hemodialysis (CVVHD), HCO3- in post membrane blood will exchange interaction. The exchange results of HCO3- determine the impact of CVVHD on the CO2 removal efficiency of the ECCO2R combined CRRT system. This study aims to investigate the effects of CVVH and CVVHD on in vitro CO2 clearance efficiency.
Looking for future studies?
Notify Me18 year–85 year
All sexes
Observational
Nanjing, Jiangsu, 210009, China
In the design of extracorporeal carbon dioxide removal (ECCO2R) combined with continuous renal replacement therapy (CRRT) equipment, the pre dilution replacement solution for continuous veno-venous hemofiltration (CVVH) enters the extracorporeal bloodstream upstream of the membrane lung, so the HCO3- concentration in the pre membrane lung blood is diluted by the replacement solution. HCO3- and dissolved CO2 are in a dynamic and rapid equilibrium state, and a decrease in HCO3- leads to a decrease in PCO2. On the other hand, in continuous veno-venous hemodialysis (CVVHD), the blood after the membrane needs to undergo material exchange with the dialysate through a filter. The dialysate is usually HCO3- with normal plasma concentration, so HCO3- in post membrane blood will exchange interaction. The exchange results of HCO3- determine the impact of CVVHD on the CO2 removal efficiency of the ECCO2R combined CRRT system. This study aims to investigate the effects of CVVH and CVVHD on in vitro CO2 clearance efficiency.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
continuous veno-venous hemofiltration or continuous veno-venous hemodialysis set-up: predilution 30 ml/kg/h, dialysate dose 30 ml/kg/h, ultrafiltration rate was equal to the total rate of substitution fluid infusion
Time frame: after 30min of continuous renal replacement therapy
extracorporeal CO2 removal rate was calculated as the difference of total CO2 content before membrane lung and after hemofilter, and normalized to an inlet PCO2 of 45 mmHg
Southeast University, China
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06944093
Respiratory Failure With Hypercapnia
Erzurum, Turkey (Türkiye)
View Trial DetailsNCT06114667
Acid-Base Imbalance, Acidosis
Montpellier, Montepllier, France
View Trial DetailsNCT06988111
Postoperative Respiratory Complication, Respiration Disorders
Asyut, Assuit, Egypt
View Trial DetailsNCT06970041
Hypoventilation, Pathological Conditions, Signs and Symptoms
View Trial Details