The Effect of 10 cm Catheter Mount on Malfunction of Capnometers Under the Conditions of Water

Capnometer has been used to assess the success of intubation and the quality of cardiopulmonary resuscitation. However, capnometers can malfunction under the conditions such as pulmonary edema and hemorrhage due to the vulnerability of capnometers for water. Investigators hypothesized that the use of 10cm catheter mount will reduce the malfunction of capnometers under the conditions of water.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–60 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Hanyang University

Seoul, 133-792, South Korea

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Healthy adults without underlying disease

Exclusion criteria

  • Lung and heart disease
  • Drinking carbonated beverages

Treatment and study plan

No-DCCM

Device

Other names: Proximal capnometer of DCCM, Abbreviation: DCCM, direct connect catheter mount

DCCM

Device

Other names: Distal capnometer of DCCM, Abbreviation: DCCM, direct connect catheter mount

Water 5ml in tracheal tube

Other

Water 10ml in tracheal tube

Other

Primary outcomes

  1. the malfunction rate of capnograph

    Time frame: 1 day

    The malfunction of capnograph is defined that the ETCO2 level is out-of the measured range of each patient under water-free conditions. The calculation formula of malfunction rate is as follows: malfunctioning time of capnograph / total experiment time (2min).

    • Water conditions in tracheal tube
    • Water 5ml
    • Water 10ml
    • Capnometers
    • Mainstream capnometer
    • Microstream capnograph
    • Comparison group
    • No-DCCM group (proximal capnometer of DCCM)
    • DCCM group (distal capnometer of DCCM)

    Abbreviation : DCCM, Direct connect catheter mount

Secondary outcomes

  1. The cumulative survival rate of capnometers under the conditions of water

    Time frame: 1 day

    • Kaplan-Meier analysis

    1-1. Water conditions in tracheal tube

    • Water 5ml
    • Water 10ml

    1-2. Capnometers

    • Mainstream capnometer
    • Microstream capnograph

    1-3. Comparison group

    • No-DCCM group (proximal capnometer of DCCM)
    • DCCM group (distal capnometer of DCCM)

    Abbreviation : DCCM, Direct connect catheter mount

  2. The consistency of ETCO2 between No-DCCM group and DCCM group under the water-free conditions

    Time frame: 1 day

    • Bland-Altman plots under water 0ml in tracheal tube.

    1-1. Capnometers

    • Mainstream capnometer
    • Microstream capnograph

    1-3. Comparison group

    • No-DCCM group (proximal capnometer of DCCM)
    • DCCM group (distal capnometer of DCCM)
    • The intraclass correlation coefficient (ICC) under water 0ml in tracheal tube

    2-1. Capnometers

    • Mainstream capnometer
    • Microstream capnograph

    2-2. Comparison group

    • No-DCCM group (proximal capnometer of DCCM)
    • DCCM group (distal capnometer of DCCM)

Sponsors and collaborators

Lead sponsor

Hanyang University

Other

Registry information

Important dates

Study start
2014
Primary completion
2014
Study completion
2014
First posted
Feb 26, 2014
Registry last updated
Mar 18, 2014

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.