NCT Number: NCT02692690

TENS in Optimizer(R) Patients

Systemic skeletal muscle atrophy (cachexia) is a typical side effect of chronic heart failure.

Patients with cachexia can benefit from transcutaneous electrical neuromuscular stimulation (TENS). It is applied to strengthen the musculature and improve exercise tolerance and quality of life.

It is unclear if application of TENS can cause electromagnetic interference (EMI) in patients with Optimizer® pacemaker systems for improvement of heart failure.

This pilot safety study enrolled 6 patients with chronic heart failure and reduced left ventricular ejection fraction. All patients had an Optimizer® pacemaker system implanted for a minimum of 6 months, no patients had any ventricular arrhythmic episode 3 months prior to the study, and all were in clinically stable condition and willing to participate in this study.

Neuromuscular stimulation was administered via TENS in 4 different modes (HF, LF, Burst, MF) on the right and left musculus trapezius and on both musculi quadriceps femoris. All patients were tested for EMI without or with cardiac contractility modulation (CCM) delivery of the Optimizer®-system respectively.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–99 year

Sex eligibility

Male

Study type

Interventional

Phase

Phase 4

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • >18 years, with implantation of the Optimizer ® >6 months ago, clinically stable condition and without any ventricular arrhythmic episode in the last 3 months.

Exclusion criteria

  • Local swelling or edema at the site of stimulation, intolerance to electrical stimulation, or presence of other metallic implants at the site of stimulation

Treatment and study plan

Stimulette r4x

Device

Primary outcomes

  1. The patients with CCM-system are assessed for EMI during TENS-application

    Time frame: 6 months

    Electromagnetic interference can cause artificial sensing in pacemaker or ICD system. These artefacts may have the effect of suppressing the pacemaker pulses sensing of tachycardia in ICD.

    The patients will be exposed to electrical stimulations, which will be of different forces on the neck and the upper thigh with r4x. These impulses and the effects will be noted down. It will be investigated if there is an interference with the optimizer ® depending on the different sites of the electrical stimulations and the different impulse forces.

    Many patients with an Optimizer ® have simultaneously a pacemaker - or ICD implant. A special extra monitor will note if there are probable interferences with these pacemaker- or ICD implants.

Sponsors and collaborators

Lead sponsor

Medical University of Vienna

Other

Registry information

Official study title

Electromagnetic Interference by Transcutaneous Neuromuscular Electrical Stimulation in Patients With Optimizer(R) - a Pilot Safety Study

Important dates

Study start
2015
Primary completion
2015
Study completion
2015
First posted
Feb 26, 2016
Registry last updated
Feb 26, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.