Skip to main content
OpenTrials
Completed

NCT Number: NCT02387775

Temperature Control With an Esophageal Cooling Device in Post Cardiac Arrest Patients

This is a prospective, interventional study aiming to assess the effectiveness of the Esophageal Cooling Device (ECD) as a temperature control modality in post cardiac arrest patients. In addition, observed adverse events during ECD use, ease-of-use, nurse satisfaction and patient outcomes will be examined.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University Hospital, London Health Sciences Centre, University of Western Ontario, London, Ontario, Canada

Loading trial locations.

About this study

Temperature control in comatose survivors of cardiac arrest, is a critical aspect of these patients' care. In this context, mild temperature reduction for 24 hours post cardiac arrest has been shown to improve neurological outcomes.

Hypothermia is commonly induced using a combination of internal and external cooling modalities. Internal cooling modalities include intravenous administration of cold crystalloids and intravascular cooling catheters. External or body surface cooling can be achieved using cooling blankets, adhesive pads, and ice packs. Each of these methods however, has its limitations.

Administration of intravenous refrigerated crystalloid (4°C) boluses is a simple, effective and widely available method of hypothermia induction. Lack of precise temperature control and the potential for pulmonary edema however, make this modality unsuitable for the maintenance phase of hypothermia. Surface cooling methods, such as cooling blankets and ice packs, often cause shivering, skin breakdown, and in obese patients may be ineffective. Intravascular cooling catheters require the time of a physician for insertion and bear the potential risks of line infection and deep venous thrombosis. Searching for a temperature control device that overcomes the limitations, inefficiencies and inconveniences of existing modalities is therefore strongly desired.

The ideal temperature control modality should be effective, predictable and easy to use. An esophageal cooling device has recently become available which in theory may, may possess some of these attributes. To date, there have been no published studies examining this device's efficacy. In addition, no study has documented any adverse events during it's use, or evaluated it's ease-of-use at the bedside.

This is a prospective, interventional study aiming to assess the effectiveness of the Esophageal Cooling Device (ECD) as a temperature control modality in comatose survivors of cardiac arrest (the E-Chill trial). In addition, observed adverse events during ECD use, ease-of-use, nurse satisfaction and patient outcomes will be examined.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients admitted to the intensive care unit after resuscitation from an out-of-hospital cardiac arrest.
  • Sustained return of spontaneous circulation (no chest compressions required for >20 minutes)
  • Not obeying commands to verbal instruction.
  • On mechanical ventilation via endotracheal tube or tracheotomy.

Exclusion criteria

  • Known esophageal deformity (known esophageal varices, history of esophagectomy, previous swallowing disorders, achalasia, etc.) or evidence of esophageal trauma.
  • Known ingestion of acidic or caustic poisons within the prior 24 hours.
  • Patients with less than 40 kg of body mass.
  • Patients known to be pregnant.
  • Uncontrolled coagulopathy.
  • Documented intracranial hemorrhage.
  • Patients presenting to the intensive care unit with a temperature < 35°C.
  • Time from collapse to return of spontaneous circulation > 45 minutes or unknown.
  • Time from collapse to starting chest compressions > 10 minutes or unknown.
  • Time from collapse to assessment for enrollment > 12 hours.
  • Patients thought to have a very low chance of surviving the next 48 hours as assessed by the research team.

Treatment and study plan

Esophageal temperature control

Device

Insertion of the Esophageal Cooling Device (ECD) will take place as described in the product monograph. The ECD will be connected to the heat exchange unit Blanketrol Hyper-Hypothermia System® (Cincinnati Sub-Zero, Cincinnati, OH) and will be used for temperature control for a total of 36 hours. A patient target of 35°C will be set on the heat exchange unit. Once target is attained, it will be maintained for 24 hours. Slow rewarming will then be achieved by increasing the input target by 0.2°C per hour, until a target of 36.6°C is reached.

All temperature data (device, input target, and patient bladder temperature) will be captured automatically using Data Export Software® (Cincinnati Sub-Zero, Cincinnati, OH) on an external laptop connected to the external heat exchange unit.

Other names: Esophageal Cooling Device®

Primary outcomes

  1. Percent of time maintained within 0.5°C range of the 35°C input target, in the hypothermia maintenance phase.

    Time frame: Within 36 hours of ECD insertion.

    Once patient temperature is within 0.5°C from the 35°C target, the maintenance phase will commence. The amount of time patient temperature is maintained within ± 0.5 from 35°C, as a percentage of the total hypothermia maintenance time (24 hours) will be examined. Confidence intervals for the results will be reported.

Secondary outcomes

  1. Cooling rate in hypothermia induction phase.

    Time frame: Within 10 hours of ECD insertion.

    Average rate of temperature drop during the hypothermia induction phase. The induction phase ends once the patient temperature is within ± 0.5 from the 35°C target. Confidence intervals for the results will be reported.

  2. Percent of time patient temperature is within 0.5°C range of the input target, 30 minutes after each target elevation in the rewarming phase.

    Time frame: Within 36 hours of ECD insertion.

    Rewarming time will be considered to start, 30 minutes after the input target is increased from 35 to 35.2°C. Since patient temperature is likely to change slowly after that change, the recorded temperature taken 30 minutes after the target change will be used. Therefore if the target temperature is changed to 35.2°C at noon, the patient temperature recorded at 12:30 will be used. Whether or not the patient is within ± 0.5 from the 35.2°C target will be calculated using the patient temperatures recorded at 12:30 till the next change in target temperature. And for each subsequent elevation in target, this same rule will be applied. This calculation will be taken to reflect the degree of control over rewarming. Confidence intervals for the results will be reported.

  3. ECD insertion time.

    Time frame: Within 4 hours of enrollment.

    Time taken to insert the Esophageal Cooling Device (ECD) will be measured. This will only include time needed to insert the ECD orally and advance it to the appropriate depth. This will exclude the steps prior to insertion (measuring the appropriate depth of insertion externally, connecting the ECD to the external heat exchange unit, and lubing the tip of the ECD). Taping the ECD in place will be excluded from this time recording as well.

  4. Adverse events during and after ECD use.

    Time frame: Within 4 weeks of enrollment or until hospital discharge whichever comes first.

    Recorded events during or after ECD use: cardiac arrhythmias, bradycardias, myocardial infarction/re-infarction, ventilator associated pneumonia, gastrointestinal bleeding (any), esophageal injury, dysphagia, odynophagia, aspiration pneumonia, non-aspiration pneumonia, tube feed intolerance, reflux and evidence of esophagitis.

  5. Ease-of-use.

    Time frame: Within 40 hours of enrollment.

    Nurses taking care of patients with the device will be asked to rate the following:

    I. Observed ease of insertion of device 1 = poor, 3 = average, 5 = excellent. II. General ease-of-use and handling: 1 = poor, 3 = average, 5 = excellent. III. How the ECD compares to our standard methods of inducing hypothermia: 1 = inferior, 3 = equivalent, 5 = superior.

    IV. Since ECD insertion, have any other cooling modality been used to cool or warm the patient? If yes, please specify.

  6. Patient neurological outcome as defined by the Cerebral Performance Category.

    Time frame: Within 4 weeks of enrollement or until hospital discharge whichever comes earlier.

    Patient neurological outcome will be dichotomized into good or bad. Good outcome will be equivalent to CPC scores of 1 & 2 (where the patient is independent), and bad outcome will be equivalent to CPC scores of 3, 4 & 5 (where the patient is either dependent or dead).

    CPC Scale:

    • Functioning normally and independent, possibly with a minor disability.
    • Moderately disabled, still independent.
    • Conscious but with a severe disability, dependent.
    • Unconscious (comatose or in a persistent vegetative state).
    • Brain dead or dead by traditional criteria.
  7. Need for other cooling or warming modalities during ECD use.

    Time frame: Within 36 hours of ECD insertion.

    The requirement for any of the following cooling modalities during ECD use will be documented.

    • Intravenous refrigerated saline. Data collected: times of use, volumes administered, and reasons for use as per bedside nurse.
    • Cooling or warming blanket use. Data collected: times of use, duration of application, and reason for use as per bedside nurse.
    • Ice-pack application. Data collected: time of use, body location(s) applied, duration of application, and reason for application as per bedside nurse.
    • Body cavity cold fluid irrigation e.g. bladder, peritoneal... etc. Data collected: time of use, body cavity site, reason for use as per bedside nurse.

Other outcomes

  1. Shivering during ECD use.

    Time frame: Within 36 hours of ECD insertion.

    The development of shivering in patients not receiving neuromuscular blockade during ECD use will be examined. This will not be evaluated however, in patients receiving neuromuscular blockade as part of their hypothermia protocol.

Sponsors and collaborators

Lead sponsor

London Health Sciences Centre Research Institute OR Lawson Research Institute of St. Joseph's

Other

Collaborators

  • Western University, Canada

Registry information

Official study title

Temperature Control With an Esophageal Cooling Device: A Feasibility Study in Post Cardiac Arrest Patients

Acronym: E-Chill

Important dates

Study start
2015
Primary completion
2018
Study completion
2018
First posted
Mar 13, 2015
Registry last updated
Jul 30, 2018

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.