Duke CTSI Translational Population Health Office
Kannapolis, North Carolina, 28081, United States
NCT Number: NCT04713748
Purpose and objective: The purpose of this pilot study is participant user acceptance testing of the Tasso-SST OnDemand device to collect blood for antibody testing. The study will also compare antibody testing results from blood collected using the Tasso-SST OnDemand device compared to blood collected by phlebotomy.
Study activities and population group: All participants enrolled in the MURDOCK C3PI Study (Pro Pro00105703), and actively participating in the testing cohort (n=300), will be emailed about the study. One or more reminder emails may be sent. The study seeks to enroll up to 100 participants.
Participants will collect blood using the Tasso-SST OnDemand at the same visit where blood will be collected via phlebotomy for the MURDOCK C3PI study. Blood samples will be spun and processed for serum, which will be shipped frozen to analytical lab at Duke. Participants will complete a short survey about their experience using the Tasso-SST OnDemand device.
Data analysis and risk/safety issues: The study team will evaluate correlation between results from blood collected by the two methods and qualitatively exam participant survey results. There are no risk/safety issues. Risk of reaction at the collection site is similar to that of phlebotomy and is described in the electronic consent form.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Kannapolis, North Carolina, 28081, United States
Purpose: The purpose of this pilot study is participant user acceptance testing of the Tasso-SST OnDemand device to collect blood for antibody testing. The study will also compare antibody testing results from blood collected using the Tasso-SST On-Demand device compared to blood collected by phlebotomy.
Design and Procedure: Participants will collect a small blood sample using the Tasso-SST OnDemand device during a serology blood draw visit already scheduled for the MURDOCK C3PI Study (Pro00105703). Participants will be provided a kit that includes written, step-by-step instructions that include pictures. Participants will also have the option to watch a brief video on using the device, available online, prior to collection. The Tasso-SST OnDemand device collects whole capillary blood, at a minimum volume of 80ul.
A standard blood draw will also be done and both blood samples will be spun and processed for serum. Serum will be frozen and shipped to the lab at Duke for antibody testing. Antibody results will be compared between the two methods.
Participants will be sent a brief survey to learn about their experience using the Tasso-SST OnDemand device.
Selection of Subject:
Subject Recruitment: All participants meeting eligibility criteria will be emailed about the study. One or more reminder emails may be sent about the study. The study seeks to enroll up to 100 participants.
The Tasso-SST OnDemand device sticks to the skin with a light adhesive. When the button is pressed, a vacuum forms and a lancet pricks the surface of the skin. The vacuum draws blood out of the capillaries and into a tube attached to the bottom of the device. The Investigator Brochure is attached to this submission.
Risk and Benefit Assessment: The Tasso-SST OnDemand device uses an adhesive - the device should be removed immediately if there are any signs of an allergic reaction. There is risk of bruising or soreness at the site where blood is drawn. There is also a slight possibility of infection, and a rare risk of fainting.
Participants can minimize risk by closely following the instructions provided with the kits.
Trained staff will be present should the participant have any concerns or issues.
The study does not have a direct benefit to participants. The information learned could lead to future use of the Tasso-SST OnDemand device to collect blood safely at home.
Data Analysis and Data Safety Monitoring: The study team will evaluate correlation between results from blood collected by the two methods in this pilot study and qualitatively exam participant survey results. Summary statistics will be generated for survey responses, and qualitative data will be reviewed.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Participants will collect a small blood sample using the Tasso-SST OnDemand device during a serology blood draw visit already scheduled for the MURDOCK C3PI Study (Pro00105703). Participants will be provided a kit that includes written, step-by-step instructions that include pictures. Participants will also have the option to watch a brief video on using the device, available online, prior to collection. Both blood samples will be spun and processed for serum. Serum will be frozen and shipped to the lab at Duke for antibody testing. Antibody results will be compared between the two methods.
Participants will have a small blood sample collected via regular venipuncture.
Time frame: up to 17 days after onset of symptoms
Sensitivity is defined as the number detected as positive on the new screen (Tasso) divided by the number defined as positive under the 'gold standard' multiplied by 100%. This is presented as a comparison to testing of blood samples obtained by routine phlebotomy as the gold standard (Abbott Alinity nucleocapsid IgG antibody assay drawn by a phlebotomist).
Time frame: up to 17 days after onset of symptoms
Interclass correlation (ICC) uses the 2 measures (Tasso & SST), measured on the same subjects. ICC is defined as the Variance between Persons (Var-P) divided by the Variance within Persons (error) Var-E.
ICC is defined as Var-P/(Var-P + Var-E). confidence intervals are easily defined by incorporating sample size.
Presented as the interclass correlation coefficient between TASSO-SST collected and routine phlebotomy sample results.
Time frame: up to 17 days after onset of symptoms
Bias is measured using the Bland-Altman method. Here, the difference the values by the 2 measures (Phlebotomy-Tasso) (difference) is compared to the average of Phlebotomy & Tasso (average). Bias is defined as differences which vary from 0, and is measured by (1) assessing if the average difference is non-zero, and (2) if the difference is systematically more or less close to 0 as the average increases. Presented as the regression slope.
Duke University
Other
Pilot Study to Compare Serology Results From Blood Collected Using Tasso-SST OnDemand Device Compared to Venipuncture
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.