NCT Number: NCT02358408

System Accuracy Evaluation of Karajishi Contour and Karajishi TS Blood Glucose Monitoring Systems Following ISO 15197:2013

The study consists of 2 parts:

The objective of part 1 of this study is the performance of a short evaluation of system accuracy for 25 subjects in order to familiarize the study personnel with the study BGMS to be used during the planned evaluation of system accuracy (see IDT-1416-BL (Part 2)), ensure that the Karajishi meters are fully functioning, assure and document the accuracy of the comparison method, evaluate and validate the processes and procedures of the study methodology. The objective of part 2 of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated.

This study will be performed for Karajishi Contour and Karajishi Contour TS (manufacturer: Bayer Healthcare Diabetes Care) with 3 reagent system lots for each BGMS.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Ulm, 89081, Germany

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Male or female, with type 1 diabetes, type 2 diabetes or no diabetes
  • For provoked blood glucose excursions due to insulin dose adjustment: Male or female with type 1 diabetes or type 2 diabetes and intensified insulin therapy or insulin pump therapy.
  • Minimum age of 18 years
  • Signed informed consent form
  • Legally competent and capable to understand character, meaning and consequences of the study

Exclusion criteria

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Physical or mental constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Incapability of giving informed consent
  • < 18 years
  • Legally incompetent
  • Accommodated in an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent from investigator or sponsor
  • For provoked blood glucose excursions 50 - 80 mg/dl:
  • Coronary heart disease
  • Condition after myocardial infarction
  • Cerebral incidence
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

Treatment and study plan

Blood glucose monitoring systems for self-testing

Device

Other names: Karajishi Contour, Karajishi Contour TS

Primary outcomes

  1. Evaluation of System Accuracy according to ISO 15197 (see description)

    Time frame: For each subject, the experimental phase has an expected duration of up to 6 hours

    System accuracy criteria:

    • Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria:
    • Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
    • Each lot shall pass acceptability criterion A.
    • Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
    • Applying ISO 15197:2003 accuracy criteria:
    • 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 75 mg/dl (4.2 mmol/l) or within ± 20 % at glucose concentrations ≥ 75 mg/dl (4.2 mmol/l).

Sponsors and collaborators

Lead sponsor

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Other

Collaborators

  • Bayer

Registry information

Important dates

Study start
2015
Primary completion
2015
Study completion
2015
First posted
Feb 9, 2015
Registry last updated
Jan 28, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.