Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Ulm, 89081, Germany
NCT Number: NCT02531386
The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. In this study, system accuracy evaluation will be performed with CE-marked BGMS (BGMS = Blood Glucose Monitoring System) common among insulin pump users - Accu-Chek Aviva Nano and Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan) Contour Next Link 2.4 (Bayer Healthcare Diabetes Care) and Accu-Chek Mobile with three different reagent system lots for each BGMS.
The study will be performed in 2 parts. Each part will comprise testing of 3 BGMS.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Ulm, 89081, Germany
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Other names: Accu-Chek Aviva Nano, Accu-Chek Performa Nano, FreeStyle Lite, OneTouch Verio IQ, Contour Next Link 2.4, Accu-Chek Mobile
Time frame: For each subject, the experimental phase has an expected duration of up to 6 hours
System accuracy criteria
Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria:
Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
Each lot shall pass acceptability criterion A.
Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
Applying ISO 15197:2003 accuracy criteria:
95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 75 mg/dl (4.2 mmol/l) or within ± 20 % at glucose concentrations ≥ 75 mg/dl (4.2 mmol/l).
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Other
System Accuracy Evaluation of Accu-Chek Aviva Nano (Roche Diagnostics), Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan), Contour Next Link 2.4 (Bayer Consumer Care), One Additional CE-marked Blood Glucose Monitoring System Following ISO 15197:2013
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details