geko™
DeviceAssess if cardiac pacemakers detect geko™ electrical output pulse.
NCT Number: NCT04391257
The gekoTM device is a small disposable battery powered device that attaches to the lower leg near the knee and stimulates the common peroneal nerve, causing muscle contraction of the lower leg. This contraction increases blood flow and the device is used to treat several conditions including deep vein thrombosis and venous leg ulcers.
There is evidence that some powered muscle stimulators and related devices can affect Cardiac Demand Pacemakers. This study is to ascertain if there is any such interaction between the gekoTM device and permanently implanted cardiac pacemakers.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
University Hospital of Wales, Cardiff, United Kingdom
CFR.890.5850 requires that the labelling for powered muscle stimulators includes the contraindication 'Powered muscle stimulators should not be used on patients with cardiac demand pacemakers'.
Further, IEC 60601-2-10:2012 clause 210.7.9.2.101d, requires that the instructions for use provide: 'Advice that a PATIENT with an implanted electronic device (for example a cardiac pacemaker) should not be subjected to stimulation unless specialist medical opinion has first been obtained'. However, no details on what criteria the specialist should apply.
There is evidence that some powered muscle stimulators, and related devices can affect Cardiac Demand Pacemakers(1,2). Traditional powered muscle stimulators use pulse trains at a frequency of at least 30Hz(3) to elicit muscle contractions. The electrodes for these devices may be placed at number of sites on the body depending on the application, including the legs, arms and torso, often with well-spaced electrodes that may be attached to different limbs, sending electrical impulses through the torso.
The geko™ range of devices, however operate by delivering a single pulse at a frequency of 1Hz with electrodes placed topically, close together, at the knee. The nature of the stimulation pulse, and its location make it extremely unlikely that the geko™ devices could interfere with cardiac demand pacemakers. However, there is no direct evidence to support this assumption. This trial will determine if there are any adverse interactions between the geko™ and cardiac pacemakers.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Assess if cardiac pacemakers detect geko™ electrical output pulse.
Time frame: One day - single follow-up.
Interference with pacemaker function.
Time frame: One day-single follow-up.
Incidence of adverse and serious adverse events.
Firstkind Ltd
Industry
A Prospective Trial to Determine if it is Safe to Use the gekoTM Neuromuscular Electrostimulation Device in a Pacemaker Population.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06750692
Brain Diseases, Cardiovascular Diseases
Afyonkarahisar, Turkey (Türkiye)
View Trial DetailsNCT04992195
Alzheimer Disease, Arterial Thromboembolism
Hong Kong
View Trial DetailsNCT07736573
Brain Diseases, Cardiovascular Diseases
Beijing, Beijing Municipality, China
View Trial DetailsNCT07237633
Brain Diseases, Cardiovascular Diseases
Singapore
View Trial Details