Staple Line Reinforcement
DeviceENDOPATH ECHELON Staple Line Reinforcement is used for reinforcement of staple lines during lung resection and bariatric surgical procedures, according to instructions for use.
NCT Number: NCT04544865
This prospective, single-arm, multi-center evaluation will collect clinical data in a post-market setting. The two types of procedures studied will be gastric and thoracic. Investigators will perform each procedure using the device in compliance with their standard surgical approach and the ECHELON ENDOPATH Staple Line Reinforcement instructions for use.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Indiana University Health, Indianapolis, Indiana, United States
The primary objective of this study is to prospectively generate clinical data on device-related adverse events in a post-market setting using ECHELON ENDOPATH Staple Line Reinforcement per its instructions for use. There will be no blinding in this study. An interim analysis is planned for sample-size re-estimation.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
ENDOPATH ECHELON Staple Line Reinforcement is used for reinforcement of staple lines during lung resection and bariatric surgical procedures, according to instructions for use.
Time frame: From Day 0 (day of surgery) through 70 days post-surgery
Number of participants who experienced specific device-related AEs post gastric resection procedures was reported. Specific device-related AEs were bleeding (defined as occurrence of post-surgery blood transfusion deemed related to bleeding at the staple line or return to surgery room before 70-day post-surgery follow-up visit due to bleeding deemed related to the staple line), leak (defined as occurrence of intra-surgery or post-surgery gastrointestinal leak related to the staple line as documented intra-surgery, by clinical exam, or radiographically), and stricture (defined as occurrence of stricture documented radiographically or by endoscopy along the staple line).
Time frame: From Day 0 (day of surgery) through 70 days post-surgery
Number of participants who experienced specific device-related AEs post lung resection procedures was reported. Specific device related AEs were prolonged air leak (occurrence of intra-surgery or post-surgery lung leak related to the staple line as documented intra-surgery, by clinical exam, or radiographically) deemed related to the staple line (greater than post-surgery Day 7), and empyema (defined as purulent fluid collection in the pleural space documented radiographically, excluding chronic empyema).
Time frame: Day 0 (Intra-surgery)
Number of study devices replaced during surgery due to slipping or bunching or not properly loaded onto stapler cartridge, Other - preloaded wrong color cartridge, Other- SLR added before cartridge, Other- dropped on the ground was reported.
Time frame: Day 0 (Intra-surgery)
Investigators were asked to answer a non-validated device questionnaire related to his/her experience using the ECHELON SLR. Questionnaire included following questions "experienced less manipulation or movement compared to previous device, experienced greater confidence the device will deliver best result compared to previous device, experienced greater confidence in simplicity of set up compared to previous device, experienced less frustration with device compared to previous device, experienced less waste compared to previous device, Likely to recommend to a colleague. Each question response was ranged from 1 to 5 where 1=strongly disagree, 2=slightly disagree,3=neutral, 4=slightly agree, 5= strongly agree, where the higher response signified good experience on the device. In this outcome measure, device questionnaire responses which were provided by selected investigators after the site's first, third, and 5th procedure were reported.
Time frame: Day 0 (Intra-surgery)
Investigators were asked to answer a non-validated device questionnaire related to his/her experience using the ECHELON SLR. Questionnaire included following questions "satisfaction of operative flow compared to previous device and satisfaction of manipulation and repositioning compared to previous device" and each question response was ranged from 1 to 5 where 1=very dissatisfied, 2=dissatisfied, 3=neither dissatisfied nor satisfied (NDNS), 4=satisfied, 5=very satisfied, where higher response signified good satisfaction with the device. In this outcome measure, device questionnaire responses which were provided by selected investigators after the site's first, third, and fifth procedure were reported.
Ethicon Endo-Surgery
Industry
A Prospective, Single-Arm, Multi-Center Study of the ECHELON ENDOPATH(TM) Staple Line Reinforcement Device in Gastric and Lung Resection Procedures
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.