Bayer HealthCare LLC, Diabetes Care
Mishawaka, Indiana, 46544, United States
NCT Number: NCT01466075
The purpose of this study is to demonstrate that untrained subjects who have diabetes or pre-diabetes can operate the investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Mishawaka, Indiana, 46544, United States
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and Alternative Site Testing (AST) of the palm using the Apollo Evolution meter and an investigational sensor. Study staff test subject venous blood and all BG results are compared to a reference laboratory glucose method. Untrained subjects complete basic tasks using the User Guide and provide feedback.
Time frame: 1 hour
Untrained subjects with diabetes self-test fingerstick blood using the Apollo Evolution Investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) or +/- 20% (>=75mg/dL YSI capillary plasma). Site staff tested in parallel after subjects.
Time frame: 1 hour
Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) or +/- 20% (>=75mg/dL YSI capillary plasma).
Time frame: 1 hour
Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI venous plasma results are used to calculate the number of BGMS results within +/- 15mg/dL (<75mg/dL YSI venous plasma) or +/- 20% (>=75mg/dL YSI venous plasma).
Time frame: 1 hour
After reading the instructions for use, and without assistance from the study staff, subjects use the BGMS to perform basic tasks considered to be essential for the operation of the system.
Ascensia Diabetes Care
Industry
Clinical Study of the Apollo Evolution BGMS With Tatsu Blood Glucose Test Strips
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details