Study of an Investigational Glucose Meter System

The purpose of this study is to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Consumer Product Testing Co., Fairfield, New Jersey, United States

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Males and females, 18 years of age and older
  • Type 1 or type 2 diabetes
  • Able to speak, read, and understand English
  • Willing to complete all study procedures

Exclusion criteria

  • Pregnancy
  • Hemophilia or any other bleeding disorder
  • Previously participated in a study using the G3 system
  • Working for a medical laboratory, hospital or other clinical setting that involves training on and clinical use of blood glucose monitors.
  • A condition, which in the opinion of the investigator, would put the person or study conduct at risk (reason for exclusion will be clearly documented by investigator or designee).
  • Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company.

Treatment and study plan

G3 Investigational Blood Glucose Monitoring System

Device

Untrained subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick blood and AST of the palm using the G3 meter and an investigational sensor. Study staff test subject venous blood and all BG results are compared to a reference laboratory glucose method. Untrained subjects utilize some additional features of the meter using the User Guide and provide feedback.

Primary outcomes

  1. Percent of Self-Test Fingerstick Blood Glucose Results Within +/- 5to15mg/dL (<100mg/dL) or Within +/- 5to15% (>=100mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Untrained subjects with diabetes self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/- 5to15mg/dL (<100mg/dL YSI capillary plasma) or +/- 5to15% (>=100mg/dL YSI capillary plasma). Site staff tested in parallel after subjects.

Secondary outcomes

  1. Percent of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/dL) or Within +/- 20% (>=75mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Untrained subjects with diabetes self-test Alternative Site (AST) Palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results are compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results are used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) or +/- 20% (>=75mg/dL YSI capillary plasma).

  2. Percent of Venous Blood Glucose Results Within +/- 5to15mg/dL (<75mg/dL) or Within +/- 5to20% (>=75mg/dL) of Laboratory Glucose Method

    Time frame: 1 hour

    Study staff tested subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI venous plasma results are used to calculate the number of BGMS results within +/- 5to15mg/dL (<75mg/dL YSI venous plasma) or +/- 5to20% (>=75mg/dL YSI venous plasma).

  3. Number of Subjects Able to Perform Given Tasks Using Product Labeling for Instruction

    Time frame: 1 hour

    After reading the instructions for use, and without assistance from the study staff, subjects use the BGMS to utilize some of the additional features of the system. Study staff documents Yes or No 'Did the subject complete the task successfully?'

Sponsors and collaborators

Lead sponsor

Ascensia Diabetes Care

Industry

Registry information

Official study title

Performance of the G3 Blood Glucose Monitoring System With TATSU Strip

Important dates

Study start
2011
Primary completion
2011
Study completion
2011
First posted
Nov 18, 2011
Registry last updated
Feb 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.