Standardized Procedure for the Assessment of New-to-market Continuous Glucose Monitoring Systems

The purpose of this study is to assess the accuracy of two continuous glucose monitoring devices (the Dexcom G4 Platinum and Medtronic Enlite systems) in patients with type 1 diabetes mellitus.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Academic Medical Center

Amsterdam, North Holland, 1105AZ, Netherlands

About this study

The study will use two CE marked Continuous Glucose Monitoring (CGM) systems; Dexcom G4 Platinum CGM CE-marked 2012 (Dexcom, USA) and the Medtronic Paradigm Veo system with Enlite Sensor CE-marked 2011 (Medtronic, USA). Both the Dexcom G4 Platinum and the Medtronic Enlite sensors will be obtained through normal commercial channels rather than directly from the manufacturer. The two CGM sensors will be worn concomitantly by the participant during the visit to the Clinical Research Center (CRC). This visit will have a duration of 6 hours. Blood will be drawn for the determination of glucose levels. The patient will receive his usual breakfast and an increased insulin bolus (180% of the patient's calculated mealtime dose) will be administered to correct the breakfast glucose excursion with the aim of inducing a period of minor hypoglycaemia. Blood sampling will continue until the end of the admission. At the end of this CRC part, the patient will continue to wear the two sensors at home. In the case of sensor failure before the CRC session on the third day into the study, patients will be instructed to insert a new sensor per the manufacturer's instruction for use and to notify study coordinators of the event. Sensors will be worn until the end of the six day study duration. Patients will be asked to perform at least 6 fingersticks per day for blood glucose measurements with the study glucometer. The study will end on the 6th day after initial sensor insertion. Patients will return to the CRC to have the sensor removed and their CGM sensor and blood glucometer data downloaded from the devices. In case of failure of both sensors after the CRC session but before the scheduled six day study duration, patients will return to the CRC for sensor removal and for data download.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • aged 18 years or above
  • diagnosed with Type 1 diabetes mellitus at least 6 months according to the WHO definition
  • Body Mass Index (BMI) <35 kg/m²
  • willing and able to wear a CGM device for the duration of the study and undergo all study procedures
  • HbA1c <10%
  • signed informed consent form prior to study entry

Exclusion criteria

  • Patient is pregnant, or breast feeding during the period of the study.
  • Patient is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
  • Patient may not use acetaminophen (paracetamol) while participating in the study
  • Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the patient's safety or successful participation in the study.
  • Patient is actively enrolled in another clinical trial or took part in a study within 30 days
  • Known adrenal gland problem, pancreatic tumour, or insulinoma
  • Inability of the patient to comply with all study procedures
  • Inability of the patient to understand the patient information.
  • Patient donated blood in the last 3 months

Treatment and study plan

CGM Monitoring

Device

All patients will be fitted with the two study CGM systems. Paradigm Veo will be configured to use the CGM part only if another pump is used.

Other names: Dexcom G4 Platinum and Medtronic Enlite CGM

Primary outcomes

  1. Mean Absolute Relative Difference (MARD)

    Time frame: up to day 6 of use

    MARD will be assessed as an average of the first 6 days of wear

Secondary outcomes

  1. Accuracy of sensors per glycemic range and trial phase

    Time frame: up to day 6 of use

    Accuracy of the sensor in the hypoglycaemic (defined as a blood glucose value ≤3.9 mmol/L) and hyperglycaemic area (defined as a blood glucose value ≥10.0 mmol/L) as an average of the first six days of use. Additionally, a separate analysis will be performed to assess sensor performance and accuracy per day of sensor life.

Sponsors and collaborators

Lead sponsor

Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA)

Other

Registry information

Acronym: SPACE2

Important dates

Study start
2013
Primary completion
2013
Study completion
2013
First posted
Dec 18, 2012
Registry last updated
Dec 16, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.