General Hospital of Northern Theater Command, PLA
Shenyang, Liaoning, 123005, China
Location status: Recruiting
Location contact
Yaling Han, PhD
CONTACT
NCT Number: NCT05040074
A feasibility study to evaluate the effectiveness and safety of the SQ-Kyrin Transcatheter Edge-to-Edge Valve Repair System for the treatment of moderate-to-severe mitral regurgitation in patients with high surgical risk.
Interested in participating?
Request Info18 year and older
All sexes
Interventional
Not applicable
Shenyang, Liaoning, 123005, China
Location status: Recruiting
Yaling Han, PhD
CONTACT
This is a prospective, multicenter, single-arm observation, feasibility clinical study that plans to enroll a total of not less than 10 and not more than 30 patients, adopts the post-op immediate technical success rate as the primary endpoint, and uses the all-cause mortality at 30 days post-op, incidence of serious adverse event at 30 days post-op, device success rate at 30 days post-op, and procedural success rate at 30 days post-op as secondary endpoints to evaluate the effectiveness, safety, and feasibility of the Transcatheter Edge-to-Edge Valve Repair System of Shanghai Shenqi Medical Technology Co., Ltd. in clinical use and preliminarily evaluate the product's performance. A phasic study report (for phasic data summary and analysis) will be issued after completion of the 30-day postoperative follow-up to evaluate the effectiveness, safety, and feasibility of the device and apply for clinical studies before official registration and marketing. In the meantime, patients enrolled will continue to be followed up for 6 months and 1-5 years post-op to evaluate the long-term effect of the mitral valve repair system.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Transesophageal echocardiography-guided, implantable mitral valve clip repair system
Time frame: Immediate postoperative
Must meet all the four items:1)no procedural mortality; 2)success in delivery and retrieval of the device delivery system; 3)successful deployment and accurate positioning of the device; 4)no emergency surgery or reintervention related to the device or surgical approach
Time frame: 1 month
Death from any cause
Time frame: 1 month
If the adverse event meets any of the criteria below, it is regarded as a serious adverse event (SAE): 1) results in death; 2) is life threatening, results in illness or injury; 3) requires inpatient hospitalization or prolongation of existing hospitalization; 4) results in a persistent or significant disability/incapacity; 5) results in a congenital anomaly/birth defect, fetal death, fetal distress; 6) results in medical or surgical intervention to prevent permanent impairment to body structure or a body function; 7) an important medical event that may not result in death, be life-threatening, or require hospitalization but may be considered serious when, based upon appropriate medical judgment, may jeopardize the patient or subject and/or may require intervention to prevent one of the outcomes listed in this definition.
Time frame: 1 month
Definition of device success(must meet all the four items) :1) No procedural mortality or stroke;2) Proper deployment and positioning of the device;3) No emergency surgery or reintervention related to the device or surgical approach;4) The implanted mitral valve repair system achieves the following safety and effectiveness indicators:
Time frame: 1 month
Definition of procedural success(must meet the two items):1) Device success;2) Absence of the following severe complications: death; stroke; life-threatening hemorrhage; severe vascular complications; secondary severe organic heart diseases (e.g.: aortic dissection, left atrium/auricle rupture, left ventricular outflow tract obstruction, etc.); Grade II-III and above renal insufficiency; myocardial infarction or coronary ischemia requiring percutaneous coronary intervention (PCI) or coronary artery bypass grafting (CABG); severe hypotension, heart failure, respiratory failure requiring use of high-dose vasoactive medications/mechanical acids; any valve-related dysfunctions, including repair device detachment (single or two valve leaflets)/embolism/device-related autologous mitral valve damage/other conditions requiring second intervention.
Contact information is provided by the study sponsor or research team.
Shanghai Shenqi Medical Technology Co., Ltd
Industry
SQ-Kyrin Transcatheter Mitral Valve Repair System Feasibility Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06600191
Cardiovascular Diseases, Heart Diseases
Melbourne, Australia
View Trial DetailsNCT07450911
Cardiovascular Diseases, Heart Diseases
Tbilisi, Georgia
View Trial DetailsNCT05061004
Cardiovascular Diseases, Heart Diseases
Phoenix, Arizona, United States
View Trial DetailsNCT05496998
Cardiovascular Diseases, Deglutition Disorders
Aarhus, Denmark
View Trial Details