Element Materials Technology
Louisville, Colorado, 80027, United States
NCT Number: NCT07527000
The clinical use of pulse oximeters has reduced the frequency and necessity of invasive arterial blood sampling and has helped to improve patient safety by providing continuous information to clinicians about patients' oxygenation status. all.health is dedicated to developing and applying innovative electronic medical solutions that improve patient care with the Algo Oximeter Software, to be used in applications for monitoring oxygen saturation.
Looking for future studies?
Notify Me22 year–50 year
All sexes
Observational
Louisville, Colorado, 80027, United States
The purpose of this clinical study is to validate the SpO2 accuracy performance of the all.health SpO2 sensors over the range of 70-100% SaO2 during non-motion conditions assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the all.health pulse oximeter.
The participants will have an arterial catheter placed in the radial artery to allow for simultaneous blood samples during stable plateaus of induced hypoxic levels. An investigational device will be placed on the wrist of the side opposite the arterial catheter following the instructions for use. Additional devices may be placed on the back of the wrist with the arterial catheter using a wrist strap. The site(s) where sensors are placed will be measured. Simultaneous data collection will be set up for the device under test and control oximeter.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Novocaine),
No intervention
Time frame: April 2026-May 2026
It is required that the Accuracy Root Mean Square (ARMS) performance of the Sponsor's DUT will meet a specification of 3.0 % or better in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification. Reference CO-Oximetry will be used as the basis for comparison for the investigational device SpO2 readings.
All Health, Inc.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07284251
Device Performance
Malatya, Battalgazi, Turkey (Türkiye)
View Trial DetailsNCT01425151
Device Performance, Device Success Rate
Innsbruck, Austria
View Trial DetailsNCT07348289
Device Performance, Emergency Medicine
Park City, Utah, United States
View Trial DetailsNCT07093398
Healthy Adult Volunteers
Adelaide, South Australia, Australia
View Trial Details