NCT Number: NCT04933773

SpO2 Measurement Depending on Different Pulse Oximeter Averaging Time Settings

The aim of this study is to characterize the noise in the SpO2 signal depending on different pulse oximeter averaging time settings.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–30 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Faculty of Biomedical Engineering, Czech Technical University in Prague

Kladno, 27201, Czechia

About this study

Noise in the SpO2 signal is most often caused by artifacts, such as movement artifacts, reduced perfusion, additional light and others. These artifacts can be partially removed from the SpO2 signal by different pulse oximeter settings, specifically the sensitivity mode and averaging time. It is known that when a longer averaging time is set, the measured signal contains fewer artifacts, on the other hand there is an increased risk of missing short desaturation. The aim of this work is to characterize the noise in the SpO2 signal depending on the set averaging time of the pulse oximeter based on experimental data from healthy volunteers.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • healthy volunteers from the Czech Technical University

Exclusion criteria

  • pregnancy
  • severe cardiovascular conditions
  • injury to the upper limbs or hands that could affect the peripheral perfusion
  • diabetes
  • hypotension or hypertension

Treatment and study plan

Movement artifact - knocking

Other

60 seconds motion artifact simulation - knocking with sensors

Movement artifact - tremor

Other

60 seconds motion artifact simulation - tremor with sensors

Artifact - ambient light

Other

60 seconds artifact simulation - ambient light using sensor opening

Artifact - low perfusion

Other

60 seconds artifact simulation - reduced perfusion using a cuff

SpO2 measurement

Other

Two SpO2 measuring devices are active simultaneously on one hand during measurement.

Primary outcomes

  1. SpO2 readings with different averaging times

    Time frame: 30 minutes

    Differences in the SpO2 signal during artifacts depending on the set averaging time.

Sponsors and collaborators

Lead sponsor

Czech Technical University in Prague

Other

Registry information

Important dates

Study start
2021
Primary completion
2021
Study completion
2021
First posted
Jun 22, 2021
Registry last updated
Jan 28, 2022