Clinimark
Louisville, Colorado, 80027, United States
NCT Number: NCT04811963
To compare prototype pulse oximeter saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from healthy human subjects.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Observational
Louisville, Colorado, 80027, United States
The study utilizes a single-site, physiology laboratory (Clinimark) for non-randomized, prospective, observational studies. The goal is to enroll a minimum of ten subjects. Subjects may be of any race, ranging in pigmentation from light to dark to meet the study design requirements.
The general purpose of Invasive Controlled Desaturation Studies is to validate the SpO2 and pulse rate accuracy in comparison to reference-standard measurements of blood SaO2 by a CO-oximeter during low saturation conditions. This is achieved through paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of 60 % to 100 % SaO22 of the prototype pulse oximeter on a group of healthy adult volunteers. The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods. Arterial blood samples are periodically taken from an indwelling arterial catheter for use in the comparison.
Pulse rate accuracy will be evaluated during the same data collection period as SpO2. Pulse rate will be compared to reference ECG heart rate. In the unlikely event that the ECG monitor malfunctions, the Pulse rate reference may be taken using the average pulse rate value from the transfer standard pulse oximeters. This study is designed to support an FDA submission for pulse rate and saturation accuracy in a diverse subject population during low saturation conditions over a specified saturation range for the OxySoftN pulse oximetry system.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Following arterial cannulation, hypoxia is induced by reducing the percent of inspired oxygen the subject breathes while simultaneous measurements are recorded from the test pulse oximeters and arterial blood samples are collected at targeted levels of saturation.
Other names: Desaturation, Blood Sampling
Time frame: April 2021
To validate the proposed claims for saturation accuracy in a diverse subject population during Low Saturation over a specified saturation range. The acceptance criteria for the OxySoftN Sensor SpO2 during Low Saturation conditions across the saturation range of 60-80% when paired with the N-600x, PM1000N and Oxicable:
SpO2 OxySoftN ± 3% (Arms)
Medtronic - MITG
Industry
SpO2 Accuracy Validation of the OxySoftN Sensor Via Reference CO-Oximetry in Healthy, Well-Perfused Subjects
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05212064
Desaturation of Blood, Hypoxia
Hangzhou, Zhejiang, China
View Trial DetailsNCT04559763
Desaturation of Blood, Hypoxia
Louisville, Colorado, United States
View Trial DetailsNCT07720960
Erythropoiesis, Hypoxia
Elche, Alicante, Spain
View Trial DetailsNCT05767385
Cardiovascular Abnormalities, Cardiovascular Diseases
San Francisco, California, United States
View Trial Details