Owlet Baby Care, Inc.
Lehi, Utah, 84043, United States
NCT Number: NCT03630016
Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. There are two devices used in this investigational study: BabySat v.1.0 and Owlet Baby Care, Inc. Smart SockTM 2 (currently available over-the-counter), with specific emphasis on its pulse oximetry accuracy. The BabySat and Smart SockTM 2 are non-invasive home care devices for use with infants. The purpose of this study was to validate the SpO2 accuracy of BabySat v1.0 and Owlet Smart SockTM 2, OSS v1.1 sensors pulse oximetry, during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Interventional
Not applicable
Lehi, Utah, 84043, United States
Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. A pulse oximeter is a device that measures the oxygen saturation of arterial blood non-invasively.
The purpose of this study was to evaluate the SpO2 accuracy performance of the Owlet BabySat v1.0 and Smart SockTM 2 pulse oximetry OSS v1.1 sensors during non-motion conditions over the range of 70-100% SaO2 to arterial blood samples assessed by CO-Oximetry for a SpO2 validation.
The goal, in its entirety, was to show the SpO2 accuracy performance for the investigational device Owlet Baby Care, Inc. Smart SockTM V. 2.
It was expected that the Accuracy Root Mean Square (ARMS) performance would meet the required specification of ARMS 3% or lower in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.
No risks or adverse device effects were expected. There were no contraindications for use in the proposed study population.
The study was conducted in accordance to the code of federal regulations for non-significant risk medical device studies and applicable ISO 14155 (2nd edition 2011-02-01), applicable sections of ISO80601-2-61 (1st edition 2011-04- 01), and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013)
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors
Time frame: Acute immediate assessment of the sensor accuracy compared to CO-oximetry
The purpose of the study was to evaluate the accuracy of the blood oxygen saturation (SpO2) at rest over the range of 70-100% compared to arterial blood samples assessed by standard methods [Co-Oximetry]. A standard pulse oximeter accuracy is evaluated by the Accuracy Root Mean Square (ARMS). The results of accurate performance of the oximetry system require ARMS of 3% or less in non-motion conditions for the range of 70-100% SpO2 per standard.
Owlet Baby Care, Inc.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07720960
Erythropoiesis, Hypoxia
Elche, Alicante, Spain
View Trial DetailsNCT05767385
Cardiovascular Abnormalities, Cardiovascular Diseases
San Francisco, California, United States
View Trial DetailsNCT01542177
Digestive System Diseases, Digestive System Neoplasms
Toronto, Ontario, Canada
View Trial DetailsNCT05354648
Arterial Occlusive Diseases, Arteriosclerosis
View Trial Details