SpinChip hs-cTnI
Diagnostic TestSpinChip Platform, consisting of the SpinChip hs-cTnI test (self-contained cartridge) and SpinChip Analyzer (instrument)
NCT Number: NCT06499519
During a heart attack, the protein troponin I is released from the heart muscle into the bloodstream. Measurements of cardiac troponin in blood are used as an aid in the diagnosis of heart attack. The SpinChip hs-cTnI test is a new high-sensitive test for measuring the amount of cardiac troponin I in the bloodstream as an aid in the diagnosis of heart attack.
The purpose of this study is to evaluate the diagnostic accuracy and safety of the SpinChip hs-cTnI test relative to a clinically validated hs-cTnI method.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Aarhus Universitetshospital, Aarhus, Denmark
Cardiac troponins are widely used as a biomarker to aid in the diagnosis of acute myocardial infarction (AMI). These structural proteins are essential in regulating contraction in cardiac muscle cells, and they are sensitive and specific biochemical markers of myocardial damage.
During a heart attack, cardiac muscle cells are injured and release the cardiac marker troponin I (cTnI) into the bloodstream. High-sensitive troponin tests may detect the increase of cardiac troponin in blood within hours after the symptoms of a heart attack has started.
The SpinChip high-sensitivity cardiac troponin I (hs-cTnI) test is a new high-sensitive test for measuring troponin I in blood samples, and the analysis may be performed close to the patient (near-patient test). The results may be obtained within 10 minutes, compared to approximately 1 hour for normal laboratory analysis.
The SpinChip Platform consists of the SpinChip hs-cTnI test and the SpinChip Analyzer and may be used at the emergency department to evaluate patients presenting with symptoms of acute myocardial infarction (chest pain). The test can use blood from finger prick or venous blood samples, either as whole blood or separated into plasma
This study is a multicentre, prospective, observational, non-randomised, open clinical performance study for evaluation of the diagnostic accuracy and safety of the SpinChip hs-cTnI test as an aid in the diagnosis of AMI. Subjects presenting with acute chest discomfort or other symptoms suggestive of AMI will be recruited at the emergency departments (EDs).
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
SpinChip Platform, consisting of the SpinChip hs-cTnI test (self-contained cartridge) and SpinChip Analyzer (instrument)
Time frame: Day of admission
Calculate and compare the area under the curve (AUC) for both SpinChip hs-cTnI and clinically validated hs-cTnI method using centrally adjudicated diagnosis of AMI
Time frame: Day of admission
Calculate and compare the area under the curve (AUC) for both SpinChip hs-cTnI and clinically validated hs-cTnI method using centrally adjudicated diagnosis of AMI for the given timepoints
Time frame: Day of admission
NPV is defined as the proportion of subjects with measured cTnI concentrations below a clinical cut-off value, who are correctly diagnosed as not having MI.
The NPVs of the SpinChip hs-cTnI test will be calculated for 0, 1, 2, and 3 hours timepoints, using the overall and sex-specific 99th percentile upper reference limits as clinical cut-offs.
Time frame: Day of admission
PPV is defined as the proportion of subjects with measured cTnI concentrations above a clinical cut-off value, who are diagnosed with MI.
The PPV of the SpinChip hs-cTnI test will be calculated for cTnI concentrations obtained 0, 1, 2, and 3 hours timepoints, using the overall and sex-specific 99th percentile URL as clinical cut-offs.
Time frame: 30 days
Prognostic accuracy for risk of MI/CV death for 30 days evaluated using Kaplan-Meier plots
Cox proportional hazard model will be used to evaluate the prognostic performance of the SpinChip hs-cTnI test for outcomes MI and CV death within 30 days.
Time frame: Day of admission
Rule-in/rule-out algorithms will be calculated in accordance with the European Society of Cardiology (ESC) rapid rule-in/rule-out protocols for cTnI results obtained at admission (0h) and 1 hours, as well as 0 h and 2 hours after admission. The rule-out limit will be derived so that the sensitivity and NPV is > 99%, and rule-in limit so that the diagnostic specificity and PPV is > 70% using 60% of the study population. The derived rule-out and rule-in limits will be validated using a remaining population.
Time frame: Day of admission
Assess the safety of the SpinChip Platform, as measured by adverse events (AEs), adverse device effects (ADEs) and device deficiencies (DDs).
SpinChip Diagnostics ASA
Industry
Clinical Diagnostic Performance of the SpinChip High-sensitivity Cardiac Troponin I (SpinChip Hs-cTnI) Test
Acronym: HEAT-4
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05511649
Acute Myocardial Infarction, Cardiovascular Diseases
Kaifeng, Henan, China
View Trial DetailsNCT06463275
Acute Myocardial Infarction, Cardiovascular Diseases
Mölndal, Gothenburg, Sweden
View Trial DetailsNCT00727623
Acute Coronary Syndrome, Acute Myocardial Infarction
La Louvière, Hainaut, Belgium
View Trial DetailsNCT03677180
Acute Myocardial Infarction, Cardiogenic Shock
Birmingham, Alabama, United States
View Trial Details