Skip to main content
OpenTrials
Completed

NCT Number: NCT03811444

Simulated Clinical Use Testing On Safety Lancets (Test B)

A simulated clinical use testing on the HTL-Strefa's safety lancets

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

UL LLC, Chicago, Illinois, United States

Loading trial locations.

About this study

This study aims to evaluate the safety of the use of the safety lancets: Haemolance Plus, Prolance and MediSafe Solo in the prevention of NSI and to evaluate the user's opinion with regards to the handling characteristics of the medical devices.

The simulated use clinical study will involve healthcare professionals (HCPs) who routinely use safety lancets to collect blood samples.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

The inclusion criteria for HCP and lay person evaluators have been chosen based on FDA guidance:

  • evaluators who routinely use safety lancets to collect capillary blood samples (prevention of the learning curve artifacts)
  • evaluators will be United States (US) residents
  • evaluators can read, write, and speak English
  • evaluators are at least 18 years old
  • evaluators are able to understand and provide signed consent for the study
  • evaluators are willing to comply with the study protocol, including being willing to answer questions and complete questionnaires.
  • evaluators have no concerns about the ability to perform the simulated skin pricking.

Exclusion criteria

Individuals will be excluded if:

  • They do not routinely use safety lancets to collect capillary blood samples,
  • They cannot read, write, and speak English,
  • They or a family member have a business or consulting relationship with a pharmaceutical or medical device company, or
  • They have participated in a product evaluation

Treatment and study plan

Haemolance Plus safety lancet

Device

Haemolance Plus safety lancet type 420 is a sterile, single use medical device designed for the sampling of capillary blood from the fingertip of patients

Prolance safety lancet

Device

Prolance safety lancet type 430 is a sterile, single use medical device designed for the sampling of capillary blood from the fingertip of patients.

MediSafe Solo safety lancet

Device

MediSafe Solo safety lancet type 520 is a sterile, single use medical device designed for the sampling of capillary blood from the fingertip of patients.

Primary outcomes

  1. The true failure rate of the tested devices

    Time frame: At time of testing, up to 90 minutes

  2. Effectiveness of the safety lancet's sharps injury prevention features

    Time frame: At time of testing, up to 90 minutes

    Evaluator questionnaire using five-point response Likert scale (coded from 1 "strongly disagree" to 5 "strongly agree") to measure respondents' attitudes to a particular question or statement. Numbers assigned to Likert scale express a "greater than" relationship.

Secondary outcomes

  1. Subjective assessments regarding evaluators' interactions with the tested safety lancets

    Time frame: At time of testing, up to 90 minutes

    Evaluator questionnaire using five-point response Likert scale (coded from 1 "strongly disagree" to 5 "strongly agree") to measure respondents' attitudes to a particular question or statement. Numbers assigned to Likert scale express a "greater than" relationship.

  2. Safety of the devices based on the evaluators' assessments (five-point response scale)

    Time frame: At time of testing, up to 90 minutes

    Evaluator questionnaire using five-point response Likert scale (coded from 1 "strongly disagree" to 5 "strongly agree") to measure any and all observed needlestick injuries and device malfunctions. Numbers assigned to Likert scale express a "greater than" relationship.

  3. Various aspects of the ease of use.

    Time frame: At time of testing, up to 90 minutes

    Evaluator questionnaire using five-point response Likert scale (coded from 1 "strongly disagree" to 5 "strongly agree") to measure respondents' attitudes to a particular question or statement. Numbers assigned to Likert scale express a "greater than" relationship.

  4. Any handling, usability questions / issues associated with the device

    Time frame: At time of testing, up to 90 minutes

    Evaluator questionnaire using five-point response Likert scale (coded from 1 "strongly disagree" to 5 "strongly agree") to measure respondents' attitudes to a particular question or statement. Numbers assigned to Likert scale express a "greater than" relationship.

Sponsors and collaborators

Lead sponsor

HTL-Strefa S.A.

Industry

Registry information

Official study title

Simulated Clinical Use Testing to Evaluate Sharps Injury Prevention Features of HTL-STREFA's Safety Lancets (Test A)

Important dates

Study start
2018
Primary completion
2018
Study completion
2018
First posted
Jan 22, 2019
Registry last updated
Jan 22, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.