David Grant U.S. Airforce Medical Center
Travis Air Force Base, California, 94533, United States
NCT Number: NCT04604145
This is a study that will attempt to validate the process for detecting SARS-CoV-2 (COVID19) on a non-FDA-approved technology using self-collected saliva as the specimen. Investigators will compare self-collected saliva samples and healthcare-worker collected nasopharyngeal samples (Nasal swabs) to see if the self-collected saliva samples are similar in terms of diagnostic accuracy. Investigators will be performing this testing at the site where patients regularly go for COVID19 testing. There will be minimal risk of harm as consenting patients will only have to provide a small amount of saliva into a tube.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Travis Air Force Base, California, 94533, United States
This is a descriptive study that will attempt to validate the procedure for detecting SARS-CoV-2 on the Eppendorf Thermal Cycler PCR system using self-collected saliva as the specimen. Investigators will calculate the percent agreement and the 95% confidence intervals (ie, overall agreement, percent positive agreement, and percent negative agreement) between self-collected saliva samples (evaluated on the Eppendorf Thermal Cycler PCR system) and healthcare-worker collected nasopharyngeal samples (evaluated on either the BioFire® FilmArray® Torch® system or the Cepheid® GeneXpert® as the gold standard). Participants without usable results for both the saliva and nasopharyngeal methods will not be included in the analyses. Currently, DGMC uses both FDA Emergency Use Authorization systems interchangeably to test for SARS-CoV-2.
Using human volunteers to obtain positive and negative nasopharyngeal and saliva specimens of SARS-COV-2, this is a test validation study. It uses the Eppendorf Thermal Cycler PCR system to compare the test performance of saliva as a specimen to the standard nasopharyngeal specimen.
This protocol utilizes two different convenience samples of DGMC patients. Patients (i.e. inpatient, outpatients) with a positive COVID-19 test in the past 72 hours, and those who present for COVID-19 testing at DGMC and have negative results. Both groups of patients are willing to provide a saliva sample for COVID-19 testing within 72 hours. The populations will be comprised of active duty, DoD dependents, and retirees who are DGMC patients. This study will enroll at least 60 individuals. Only saliva samples from 30 participants with negative and 30 with positive NP results will be processed. The study will stop recruiting positive patients after obtaining the paired saliva and NP results from 30 participants who are COVID-19 positive by NP test. The study will stop actively recruiting negative patients after obtaining the paired saliva and NP results from 30 participants who are COVID-19 negative by NP test. More than 30 negative patients may be recruited in an attempt to capture 30 positive patients.
Recruitment efforts for positive patients will be conducted utilizing a HIPAA waiver to identify potential study participants, as well as, targeting patients exposed to and/or with symptoms related to COVID-19 who present for COVID-19 testing. The HIPAA Waiver will be used to identify patients (inpatient, outpatient) who have had a positive COVID-19 NP test within the past 72 hours. These patients, as well as others who will have COVID-19 testing at DGMC, will be informed of the study.
Patients interested in participation will make contact with study staff. Those eligible to participate will be given study documents (consent and HIPAA Authorization Form), that will later be reviewed and signed by the patient and study staff. Thereafter, additional information (contact, demographic), and a self-collected saliva specimen will be obtained.
To minimize contact with COVID-19 positive or potentially infected patients when signing documents and obtaining the self-collected saliva sample, staff will wear scrubs (or a lab coat), gloves, eye protection and N95 mask. Documents requiring signature will be signed by the patient while wearing a mask and maintaining a distance of at least 6 feet from study staff or from their vehicle. Consent and HIPAA Authorization Forms will be signed. The original, signed documents will be maintained by study staff and a copy will be provided to the participant. After both documents have been signed, study staff will obtain demographic (age, ethnicity, race and sex) and contact information (name, email address if future contact is desired), date of birth, and the self-collected saliva sample.
Prior to collecting the saliva sample, study staff will ensure the participant has not smoked, eaten or drank anything for the previous 10 minutes. Participants will be given a biohazard bag, label (with a 6-digit numerical code), 25 ml plastic specimen tube without transport medium inside. Then asked to fill it with approximately 2 mL of saliva. There will be a marking on the tube to ensure a sufficient sample is collected. The participant will be instructed to close the lid, place the label on the outside of the tube, then put the labeled tube into the biohazard bag. Participants will then place the bagged specimen inside a second biohazard bag held by either DGMC or a member of the study team. The double-bagged self-collected saliva sample will then be placed into the study cooler containing an ice pack. Saliva specimens will then be transported to the CIF in batch.
After obtaining the paired saliva and NP results from 30 participants who are COVID-19 negative by NP test, saliva testing will only be performed on samples collected from NP positive participants. Patients with unknown COVID-19 status, enrolled in an attempt to capture a COVID-19 positive patient, will have the medical record accessed to retrieve the NP result prior to performing saliva testing. If the NP result is positive, saliva testing will be performed.
Testing will be performed on saliva samples using reverse-transcription polymerase chain reaction (RTPCR). Natural and/or synthetic RNA will be acquired for use as a positive control. RT-PCR will target the E-gene and N-gene of SARS-CoV-2.
Once testing is completed, specimens will be aliquoted and stored at -80oC.
Saliva samples will be interpreted as negative, positive or inconclusive. Negative will be reported when there is no detection in any target; inconclusive when 1 target is detected; and positive when 2 or more targets are detected.
Study staff will access the lab accession number associated with each SARS-CoV-2 positive NP result to request a print out, from the DGMC clinical lab, of the raw data from the instrument used to perform testing. This information will be used to evaluate any discrepancies between the two testing methods.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Obtain self-collected saliva sample, perform SARS-CoV-2 testing using the Eppendorf Thermal Cycler Polymerase chain reaction (PCR) system on 30 samples associated with a positive and 30 samples associated with a negative NP sample, then compare saliva results to healthcare worker collected NP sample results evaluated on either the BioFire® FilmArray® Torch® system or the Cepheid® GeneXpert® (gold standard).
Time frame: within 1 week of SARS-CoV-2 testing
Percent positive agreement between self-collected saliva samples (evaluated on the Eppendorf Thermal Cycler PCR system) and healthcare-worker collected nasopharyngeal samples (evaluated on either the BioFire® FilmArray® Torch® system or the Cepheid® GeneXpert® as the gold standard).
Time frame: within 1 week of SARS-CoV-2 testing
Percent negative agreement between self-collected saliva samples (evaluated on the Eppendorf Thermal Cycler PCR system) and healthcare-worker collected nasopharyngeal samples (evaluated on either the BioFire® FilmArray® Torch® system or the Cepheid® GeneXpert® as the gold standard).
Time frame: within 1 week of SARS-CoV-2 testing
Percent overall agreement between self-collected saliva samples (evaluated on the Eppendorf Thermal Cycler PCR system) and healthcare-worker collected nasopharyngeal samples (evaluated on either the BioFire® FilmArray® Torch® system or the Cepheid® GeneXpert® as the gold standard).
David Grant U.S. Air Force Medical Center
Fed
Evaluation of Self-Collected Saliva Samples Without Viral Transport Media for SARS-CoV-2 Testing Via RT-PCR
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04510194
COVID-19, Coronaviridae Infections
Sylmar, California, United States
View Trial DetailsNCT05313087
Coronaviridae Infections, Coronavirus Infections
Denver, Colorado, United States
View Trial DetailsNCT04761822
Allergic Reaction, COVID-19
Tucson, Arizona, United States
View Trial DetailsNCT04895475
COVID-19, Coronaviridae Infections
Cincinnati, Ohio, United States
View Trial Details