SeCure Endovenous Laser Treatment Study

The VenaCure EVLT 400 µm Fiber Kit is currently marketed for the treatment of varicose veins. AngioDynamics, Inc., the company that manufacturers the VenaCure device, is sponsoring this study to assess the safety and effectiveness of the VenaCure EndoVenous Laser Treatment (EVLT) 400 µm Fiber Kit for treatment of incompetent perforator veins (IPVs). The VenaCure EVLT 400 µm Fiber Kit has not been previously studied by AngioDynamics for the ablation of IPVs.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Midwest Institute for Minimally Invasive Therapies, Melrose Park, Illinois, United States

Loading trial locations.

About this study

This is a single-arm, prospective, multicenter, non-blinded clinical trial. Study data will be summarized and submitted to the FDA in a premarket notification once all treated subjects have completed the 3 month visit. Longer term followup is being performed for publication purposes.

Patients diagnosed with incompetent Perforator Veins (IPV's) and meet all inclusion and none of the exclusion criteria will be eligible for this study.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

Patients are required to fulfill all the following criteria to be included in the study:

  • Is ≥ 18 years of age
  • IPV(s) to be treated have an outward flow duration of ≥ 0.5 sec immediately after manual release of manual compression
  • IPV(s) to be treated have a diameter of ≥ 3.5 mm (measured at the level of the fascia) located superior to foot and distal ankle
  • Has been diagnosed with refractory symptomatic disease (CEAP Class 4b to Class 6) attributable to the IPV to be treated
  • Has palpable pedal pulses in the study limb
  • Any pathologic superficial saphenous veins have been previously eliminated and were done so at least more than 30 days prior to the study procedure
  • Is able to ambulate
  • Is able to comprehend and have signed the Informed Consent Form (ICF) to participate in the study
  • Is willing and able to comply with the Clinical Investigation Plan and follow-up schedule

Exclusion criteria

Patients will be excluded from participation in the study if they meet any of the following:

  • Has venous insufficiency secondary to venous obstruction proximal to the intended treatment site
  • Has thrombus in the vein segment to be treated
  • Has known peripheral arterial disease
  • Has a BMI calculation (BMI = W / H2 ) ≥40kg/m2
  • Is undergoing active anticoagulant therapy for Deep VeinThrombosis or other conditions (e.g., warfarin, Q10 inhibitors or low molecular weight heparin) or has a history of Deep Vein Thrombosis within the last 6 months or hypercoagulable state.
  • Has had prior venous procedures in the study limb within the last 30 days (including but not limited to, thrombolysis / thrombectomy / stenting / ablation / phlebectomy / sclerotherapy)
  • Has undergone or is expected to undergo any major surgery within 30 days prior to or within 90 days following the study procedure
  • Has a condition, judged by the treating physician, that may jeopardize the patient's well-being and/or confound the results or the soundness of the study
  • Is pregnant or lactating at the time of the study procedure or is intending on becoming pregnant within 90 days following the study procedure
  • Is participating in another clinical study that is contraindicative to the treatment or outcomes of this investigation

Treatment and study plan

VenaCure EVLT 400 µm fiber Procedure Kit

Device

The study intervention will be conducted according to the DFU included with the VenaCure EVLT 400 µm fiber Procedure Kit.

Primary outcomes

  1. Acute Primary Ablation Success

    Time frame: 10 day

    The primary effectiveness endpoint is "Acute Primary Ablation Success" defined as complete lack of flow or IPV disappearance in the entire treated segment. Success will be measured via Duplex Ultrasound imaging performed 10 days (± 3 days) post procedure. Primary ablation success must be measured by a physician other than the physician that performed the study procedure.

Secondary outcomes

  1. Technical Success

    Time frame: Treatment

    Technical Success, defined as successful access and entry into the IPV to be ablated and the ability to deliver the intended laser energy, will be evaluated as a secondary endpoint for inclusion in device labeling.

Sponsors and collaborators

Lead sponsor

Angiodynamics, Inc.

Industry

Registry information

Official study title

Prospective Safety and Effectiveness Study: VenaCure Endovenous Laser Treatment (EVLT) 400 µm Fiber Procedure Kit for Treatment of Incompetent Perforator Veins

Acronym: SeCure

Important dates

Study start
2015
Primary completion
2018
Study completion
2019
First posted
Aug 13, 2014
Registry last updated
May 19, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.