Tampere Heart Hospital
Tampere, Finland
NCT Number: NCT05196412
The study investigates the performance pf PulseOn Arrhythmia Monitor System that includes a wrist-worn device, which combines continuous optical heartbeat interval monitoring and intermittent ECG measurement, and a data management system used by healthcare professionals for data observation, in the detection of previously undiagnosed cardiac arrhythmias, especially atrial fibrillation.
Looking for future studies?
Notify Me50 year–99 year
All sexes
Interventional
Not applicable
Tampere, Finland
The study consists of monitoring voluntary subjects with PulseOn Arrhythmia Monitoring System for the period of two weeks. The subjects are recruited mainly through newspaper advertisements. The study is executed in Tampere and Helsinki areas, Finland.
During the two week monitoring time, the subjects are instructed to wear the wrist device continuously, except when taking the device off e.g. for going to sauna or swimming, and for charging. The subjects are instructed to perform a 35-second ECG measurement (recording) in three cases:
The data collected in the study is analysed post-hoc. In the post-hoc analysis, the cardiac rhythm assessed automatically from the ECG measurements is visually confirmed by a cardiologist.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
for the main study group:
Inclusion criteria
for the active endurance athlete group:
Exclusion criteria
for both groups:
Wearing of the wrist device for a period of two weeks
Time frame: Duration of patient measurement, i.e. two weeks
Percentual improvement in the performance of detecting previously undiagnosed atrial fibrillation in the study population by the continuous PPG-based heartbeat interval monitoring when compared with relying only on scheduled and symptom -based ECG measurements. Cardiac rhythm assessed by a cardiologist from the ECG measurements that are made as a result of PPG-based arrhythmia notifications is compared with rhythm assessments made only from the scheduled and symptom-based ECG measurements. It is evaluated whether more subjects can be diagnosed having atrial fibrillation with the help of continuous PPG-based rhythm monitoring.
Time frame: Duration of patient measurement, i.e. two weeks
Data recorded by the study subjects during the two-week study time is assessed by a cardiologist and subjects are classified as having: "no observed arrhythmias", "atrial fibrillation episode(s)", "other potential arrhythmia", or "unconfident assessment" (due to the quality or lack of data). The percentage of "unconfident assessment" shall be less than 20% of the subjects.
Time frame: Duration of patient measurement, i.e. two weeks
The feedback from the subjects should show at least grade 3 in a scale of 1-5 for the "comfortability of using the wrist device".
Time frame: Duration of patient measurement, i.e. two weeks
The visual ECG-based rhythm annotation made by a cardiologist is compared with simultaneous PPG-algorithm alarm state. The incidence of notifications for other arrhythmias than atrial fibrilation is reported.
Time frame: Duration of patient measurement, i.e. two weeks
Evaluation is done by comparing the atrial fibrillation detection outcome when considering only the PPG-triggered and symptom-based ECG measurements or when considering also the scheduled ECG measurements. The outcome parameter is the number of subjects that would have been left undetected if the scheduled ECG measurements were not taken.
Time frame: At the end of the 2-week measurement period
Feedback questions in scale of 1-5 and open-ended feedback. Easiness of use of the wrist device, understandability of the user interface, suitability of the user materials, and unobtrusiveness.
Time frame: At the end of the study i.e. the end of 2022.
Feedback questions in scale of 1-5 and open-ended feedback. Easiness of use of the data management service, suitability of the user materials.
Time frame: Duration of patient measurement, i.e. two weeks
Percentage of the ECG measurements classified as inadequate quality for rhythm assessment either an ECG analysis algorithm or by a doctor.
Time frame: Duration of patient measurement, i.e. two weeks
Analysis of how well the participants will follow the study protocol and do the participants register their ECG in correlation with the wrist device notification. The outcome parameter is the percentage of scheduled and PPG-based notifications to take an ECG measurement and the variation of the percentage between the subjects.
Time frame: Duration of patient measurement, i.e. two weeks
Comparison of the time to the first detection of atrial fibrillation or other arrhythmia when considering PPG-triggered ECG measurements or when considering the scheduled ECG measurements.
Time frame: Duration of patient measurement, i.e. two weeks
The automated assessment done for the 30-second ECG recordings is compared with the visual assessment to calculate the sensitivity and specificity for each rhythm category.
PulseOn Oy
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05481359
Arrhythmias, Cardiac, Atrial Fibrillation
Athens, Georgia, United States
View Trial DetailsNCT03940066
Arrhythmias, Cardiac, Atrial Fibrillation
Santander, Spain
View Trial DetailsNCT05351775
Arrhythmias, Cardiac, Atrial Fibrillation
Turku, Finland
View Trial DetailsNCT03549806
Arrhythmias, Cardiac, Atrial Fibrillation
Stanford, California, United States
View Trial Details