NCT Number: NCT01568606

Safety of Body Composition Analysis in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)

The purpose of this study is to define the safety of using bioimpedance for analysis of body composition in heart failure patients with ICD devices. The Biospace America InBody 520 scale, using direct segmental multifrequency bioimpedance, will be utilized to assess patients' fat mass, lean muscle mass, and edema status.

Although the Biospace America InBody 520 scale is routinely used to analyze body composition in various settings including the Ahmanson-UCLA Cardiomyopathy clinic, due to theoretical concerns of safety, bioimpedance has not been routinely used in patients with ICDs. The investigators hope that this study will allow us to routinely analyze body composition in heart failure patients with ICDs, information which can be used to help guide dietary, exercise, and medical prescriptions for the investigators heart failure patients.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Ahmanson-UCLA Cardiomyopathy Center

Los Angeles, California, 90095, United States

About this study

Up to one half of heart failure (HF) patients are overweight and obese, as measured by body mass index (BMI). Comprehensive body composition assessment utilizing bioimpedance scales wuch as Biospace America InBody 520 scale, can give accurate measurements of fat mass, lean muscle mass, and edematous water weight. Body composition analysis is routinely used in many clinical settings, including the Ahmanson-UCLA Cardiomyopathy Center. However, bioimpedance scales are generally not used in patients with implantable cardioverter-defibrillators (ICDs), a group that represents a majority of our HF patient population. The safety of InBody 520 technology - direct segmental multifrequency (5, 100, 500 KHz) bioimpedance analysis - in patients with ICDs has not been formally evaluated. There is a theoretic concern that the ICD may misinterpret the applied currents for a cardiac arrhythmia and possibly cause a discharge from the ICD. However, there has been no documentation of any adverse events in patients with ICDs using the InBody 520 or other bioimpedance scales. Furthermore, thoracic bioimpedance analysis (BioZ ICG, 70 Khz) has been documented to be successfully utilized without adverse effects in patients with HF and ICDs. This study will define the safety of using the InBody 520 scale for analysis of body composition in patients with ICD devices. Subjects with HF and and ICD will have their ICD interrogated and continually monitored by an electrophysiologist before, during, and after the 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. If there is any suggestion of the ICD interpreting applied frequencies as a cardiac arrhythmia, the patient will be removed from the scale and the electrophysiologist will simultaneously temporarily deactivate their ICD device to prevent discharge. Twenty subjects (from the following four ICD companies: Medtronic, St. Jude, Guidant, and Biotronik) will participate in our study.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • HF patients of any etiology who currently have an implanted ICD

Exclusion criteria

  • patients who are non-ambulatory
  • have a physical disability making them unable to stand on the InBody
  • those who are above the height and weight maximums for the device [Height > 220cm (7'2.6''), Weight > 250kg (551lb)]
  • those who are pregnant

Treatment and study plan

Body Composition Analysis InBody Scale

Device

Patients will have their ICD interrogated using a pacemaker Programmer specifically designed from one of the 4 pacemaker manufacturers: St. Jude, Medtronic, Guidant, and Biotronik. Depending on the type of pacemaker, interrogation can be done either wirelessly or by placing a wand over the pacemaker. The ICD will be turned to a "monitor-only" mode to disable risk of ICD shock by the supervising electrophysiologist. Patient will be asked to step on the InBody 520 scale to have their body composition analyzed. Patient will continue to have their ICD actively monitored. Analysis will take approximately 30 - 50 seconds. At the end of analysis, ICD device will be returned to default settings by supervising electrophysiologist.

Other names: Body Composition Analysis InBody 520 Scale

Primary outcomes

  1. Number of Participants With Arrhythmia as Detected by ICD

    Time frame: 1 day, on day of study ICD interrogated and continually monitored before, during, and after 30-50 seconds of bioimpedance analysis

    Subjects with HF and ICD had their ICD interrogated and continually monitored by an electrophysiologist before, during, and after 30-50 seconds of bioimpedance analysis, which involves standing on the InBody 520 scale. Outcome measure was assessed during a one day visit. There is no further follow-up.

Sponsors and collaborators

Lead sponsor

University of California, Los Angeles

Other

Registry information

Official study title

Safety of Body Composition Analysis Using Bioimpedance in Heart Failure Patients With Implantable Cardioverter Defibrillators (ICDs)

Important dates

Study start
2011
Primary completion
2011
Study completion
2011
First posted
Apr 2, 2012
Registry last updated
Mar 10, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.