RELIANCE 4-FRONT™ Passive Fixation lead implantation
DeviceImplantation of transvenous defibrillation lead with passive fixation mechanism.
NCT Number: NCT01856491
The objective of this study is to gather data to establish the chronic safety, performance and effectiveness of the RELIANCE 4-FRONT™ Passive Fixation Defibrillation Leads.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
German Heart Center, Berlin, Germany
The RELIANCE 4-FRONT Passive Fixation Defibrillation Lead Post Market Clinical Follow-Up Study is a prospective, non-randomized, multi-center, single-group, post market clinical study to establish the chronic safety, performance and effectiveness of the RELIANCE 4-FRONT passive fixation defibrillation leads.
A total of 167 patients (including 10 % attrition) are required to evaluate the Primary Endpoint.
Up to 10 Investigational centers located in Europe and Israel. Lead-related Complication-Free Rate (CFR) from Implant through 3 Months Post-Implant.
Lead-related complications associated with the RELIANCE 4-FRONT passive fixation lead will count toward this endpoint.
Lead-related Complication-Free Rate from 3 Months through 15 Months Post-Implant.
Lead-related complications associated with the RELIANCE 4-FRONT passive fixation lead will count toward this endpoint.
Clinic visits will occur at:
Hypotheses testing in the RELIANCE 4-FRONT passive fixation PMCF Study will use standard statistical methodology as specified more in detail in the protocol.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Implantation of transvenous defibrillation lead with passive fixation mechanism.
Time frame: 3-months
Lead-related Complication-Free Rate (CFR) from Implant through 3 Months Post-Implant.
Time frame: 3 months through 15 months post implant
Lead-related Complication-Free Rate from 3 Months through 15 Months Post-Implant.
Time frame: 3 Months Post-Implant
Pacing Threshold at 0.5 ms pulse width at 3 Months Post-Implant. During the RELIANCE 4-FRONT PASSIVE fixation PMCF Study pacing threshold measurements are collected from RELIANCE 4-FRONT PASSIVE fixation leads in the standard manual fashion. At least 3 cardiac cycles at a given voltage level shall be obtained before stepping down to the next voltage level. A count of two non-capture beats is required at a given voltage level to declare a loss of capture (LOC) for any of these tests. The threshold is defined as one voltage level above the level where two non-captured beats are observed. Threshold tests must be taken with a pulse width of 0.5 ms.
Time frame: 3 months through 24 months post implant
Lead-related Complication-Free Rate from 3 Months through 24 Months Post-Implant
Time frame: 3 Months Post-Implant
Lead performance evaluations of the Reliance 4-FRONT PASSIVE fixation lead have been conducted in accordance with the Physician's Lead Manual. Measurements should fall within the recommended values as mentioned below:
Signal Type Amplitude Pacing Threshold Impedance Pacing/Sensing ≥ 5mV ≤1.5 V 300-1200 Ω Defibrillation ≥ 1mV N/A 20-125 Ω The required data from the implant procedure for the study lead is measured with a pacing system analyzer (PSA) to verify adequate signals. Electrical performance of the lead was verified before attaching the lead to the Pulse Generator (PG). After the PG has been implanted evaluation of the study lead using the PG has been performed. Lead measurements were required unless the testing is prohibited by a subject's condition (subject has no intrinsic rhythm). In case the measurements are variable (e.g. in patients with atrial fibrillation), the most reproducible value has been collected.
Time frame: 3 Months Post-Implant
Lead performance evaluations of the Reliance 4-FRONT PASSIVE fixation lead have been conducted in accordance with the Physician's Lead Manual. Measurements should fall within the recommended values as mentioned below:
Signal Type Amplitude Pacing Threshold Impedance Pacing/Sensing ≥ 5mV ≤1.5 V 300-1200 Ω Defibrillation ≥ 1mV N/A 20-125 Ω The required data from the implant procedure for the study lead is measured with a pacing system analyzer (PSA) to verify adequate signals. Electrical performance of the lead was verified before attaching the lead to the Pulse Generator (PG). After the PG has been implanted evaluation of the study lead using the PG has been performed. Lead measurements were required unless the testing is prohibited by a subject's condition (subject has no intrinsic rhythm). In case the measurements are variable (e.g. in patients with atrial fibrillation), the most reproducible value has been collected.
Boston Scientific Corporation
Industry
RELIANCE 4-FRONT™ Passive Fixation Defibrillation Lead Post Market Clinical Follow-Up (PMCF) Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00478933
Arrhythmia, Sinus, Arrhythmias, Cardiac
View Trial DetailsNCT00402246
Arrhythmia, Arrhythmias, Cardiac
Mobile, Alabama, United States
View Trial DetailsNCT00664807
Arrhythmia, Sinus, Arrhythmias, Cardiac
Little Rock, Arkansas, United States
View Trial DetailsNCT00180336
Arrhythmias, Cardiac, Atrial Fibrillation
Bad Nauheim, Germany
View Trial Details