Skip to main content
OpenTrials
Completed

NCT Number: NCT02168972

Safety and Performance of a GLOBAL Mapping and Ablation Device for the Treatment of Atrial Fibrillation (GLOBAL-AF)

The purpose of this study is to provide clinical data pertaining to the safety and performance of the Globe® Mapping and Ablation System for treating patients with atrial fibrillation.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Herzzentrum Leipzig GmbH, Leipzig, Saxony, Germany

Loading trial locations.

About this study

Upon successful enrolment, subjects will be scheduled for mapping and radiofrequency (RF) ablation. Cardiac CT imaging, echocardiography and baseline 12-lead ECG will be performed pre-procedure for screening and health assessment. During the procedure the Globe catheter will be delivered via standard femoral vein access and transseptal puncture. Anatomical and electrophysiological mapping of the left atrium (LA) will be followed by RF ablation to achieve pulmonary vein isolation (PVI). Additional lesions will be created as deemed appropriate by the investigator. Intracardiac electrogram mapping will be used to confirm PVI during the procedure. Echocardiography will be performed at discharge to assess cardiac function. Electrocardiograms by 12-lead ECG and 7-day Holter monitoring will be recorded at discharge, and 3, 6 and 12 months after treatment. Follow-up assessments will be performed at discharge, 7 days, and 1, 3, 6 and 12 months after treatment, to track freedom from documented AF and adverse events. Subject-reported AF symptomology and quality of life will also be assessed during follow-up.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients indicated for ablation, with a documented history of symptomatic paroxysmal atrial fibrillation
  • Between 18 and 80 years of age, inclusive

Exclusion criteria

  • Patients who have contraindications to open heart surgery
  • Patients from an Intensive Care Unit
  • Patients requiring concurrent right atrial ablation or who have had previous left atrial ablation
  • Previous cardiac procedure or known abnormalities that would interfere with device delivery, position, or treatment efficacy
  • History of a documented thromboembolic event or bleeding abnormalities
  • Contraindication to anticoagulation therapy
  • Mental impairment or other conditions, which may not allow patient to understand the nature, significance and scope of the study
  • Anatomy that would prevent safe and appropriate introduction or delivery of the Globe device into the patient

Treatment and study plan

Global mapping and ablation device

Device

During the procedure the Globe catheter will be delivered via standard femoral vein access and transseptal puncture. Anatomical and electrophysiological mapping of the LA will be followed by pulmonary vein isolation (PVI) by RF ablation. Additional lesions will be created as deemed appropriate by the investigator.

Other names: Globe® Mapping and Ablation System

Primary outcomes

  1. Rate of subjects presenting pre-defined serious adverse events occurring within 7 days of the procedure

    Time frame: 7 days

    Rate of subjects presenting with one or more of the following serious adverse events occurring within 7-days of the procedure:

    • Transient ischemic attack
    • Cerebrovascular accident
    • Major bleeding
    • Cardiac tamponade
    • Pulmonary vein stenosis
    • Pericarditis
    • Myocardial infarction
    • Diaphragmatic paralysis
    • Atrio-esophageal fistula
    • Valvular damage
    • Phrenic nerve palsy
    • Intra-procedural device complication requiring open chest or heart surgery
    • Death

Secondary outcomes

  1. Subjects presenting with adverse events

    Time frame: Up to 1 year

    • Subjects presenting with primary serious adverse events up to 1 year
    • Subjects presenting with adverse events or serious adverse events up to 1 year
  2. Acute procedural success

    Time frame: Intra-procedurally

    Acute device performance in achieving entrance block of the pulmonary veins

  3. Rate of freedom from documented atrial fibrillation

    Time frame: Between 3 months and 1 year

    Rate of freedom from electrocardiographically documented atrial fibrillation between 3 months and 1 year post-ablation

  4. Change in Quality of life and AF symptom frequency and severity scores

    Time frame: Up to 1 year

    Change of quality of life (SF-36) and AF symptom frequency and severity (Bubien et al 1993) questionnaire scores from baseline to 1 year after ablation

Other outcomes

  1. Procedure time

    Time frame: Intra-procedurally

    Procedure time

  2. Repeat ablation rate

    Time frame: Up to 1 year

    Rate of repeat ablation with a third party catheter or the Globe system

  3. Fluoroscopy time and dose area product

    Time frame: Intra-procedurally

    Fluoroscopy time and dose area product during the procedure

  4. Device functionality

    Time frame: Intra-procedurally

    Rate of procedures terminated or switched to a third-party ablation catheter due to Globe device failures

Sponsors and collaborators

Lead sponsor

Kardium Inc.

Industry

Registry information

Official study title

A Prospective, Non-randomized Clinical Study to Assess Safety and Performance of a Device for Global Mapping and Ablation of the Left Atrium for the Treatment of Atrial Fibrillation

Acronym: GLOBAL-AF

Important dates

Study start
2014
Primary completion
2017
Study completion
2019
First posted
Jun 20, 2014
Registry last updated
Oct 2, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.