Testing of the autocapture features of the device
DevicePatients receiving cardiac resynchronization therapy will have new autocapture features tested on right atrial, right ventricular and left ventricular leads.
NCT Number: NCT00746356
This study is a prospective, non-randomized, multi-center clinical trial that will assess the safety and effectiveness of investigational models of the Promote™ RF CRT-D and Current™ RF ICD devices. These new ICD and CRT-D devices have new features that allow the device to periodically check how much energy is needed to pace a patient's heart and to automatically adjust the amount of energy used to pace the heart.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Arkanasas Heart Hospital, Little Rock, Arkansas, United States
All patients enrolled in the study will have their device tested to determine if the device can automatically measure how much energy is needed to pace the patient's heart. If the test is successful, the patient's device will be programmed to turn that feature on. 3 months after enrollment, the patient will return to the clinic to have their device re-tested using an automatic test method to measure the amount of energy needed to make the patient's heart beat. A manual test method will also be used to measure the amount of energy needed to make the patient's heart beat. The intent of the study is to show the automatic method works just as well as the manual method.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Patients receiving cardiac resynchronization therapy will have new autocapture features tested on right atrial, right ventricular and left ventricular leads.
Time frame: 3 months post implant
Time frame: 3 months post implant
Atrial AutoCapture Confirm is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the atrial pacing voltage until it determines that the device is no longer capturing the atria. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the atrial pulse voltage and determines the atrial threshold by viewing the EKG.
Time frame: 3 months post implant
Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
Time frame: 3 months post implant
Right Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
Time frame: 3 months post implant
Left Ventricular AutoCapture is an automatic test that the device performs without any interaction by the physician and is one of the features being evaluated in the study. The test temporarily decreases the ventricular pacing voltage until it determines that the device is no longer capturing the ventricles. That value is reported to the physician when the device is read by the programmer. This endpoint looks at the absolute difference between this automated test and a manual test in which the physician decreases the ventricular pulse voltage and determines the ventricular threshold by viewing the EKG.
Abbott Medical Devices
Industry
A Clinical Evaluation of the Current™ Radio Frequency (RF) Implantable Cardioverter Defibrillator (ICD) Models 2215/1215 and Promote™ RF Cardiac Resynchronization Therapy Defibrillator (CRT-D) Model 3215 Pulse Generators
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial DetailsNCT05820295
Arrhythmias, Cardiac, Atrial Fibrillation
New York, United States
View Trial DetailsNCT05741658
Cardiovascular Diseases, Heart Diseases
Los Angeles, California, United States
View Trial Details