NCT Number: NCT04208282

Safety and Efficacy of 7-day Wear Infusion Set vs. Control Infusion Set.

The purpose of this study is to evaluate the Extended Wear (EW) infusion set seven day functioning in patients with Type 1 diabetes compared to the Quick-set infusion set. The study will be carried out in adults with type 1 diabetes.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Sheba Medical Center

Ramat Gan, Israel

About this study

This is a 1-center, prospective, open label randomized, crossover controlled study.

40 patients will be using 4 infusion sets (a total of 160 sets) that will serve as an exploratory pilot study to assess the SD of 7 day use for future study design.

Eligible patients meeting the enrollment and randomization criteria will be randomized into 2 arms, each arm will undergo 2 phases. In each phase a different type of infusion set will be evaluated. Each evaluation will be for4 sets (up to 28 days). The order will be determined by the randomization process:

  • Arm 1: Quick-EWIS. First phase Quickset, second phase EWIS.
  • Arm 2: EWIS-Quick. First phase EWIS, second phase Quickset.

Following screening and informed consenting to the trial, a randomized procedure will define the subject trial arm:

  • Trial Arm A: Subjects will be randomized to a group who will be using the Quickset infusion system for the Phase 1. All patients will be retrained on the use of the Quickset infusion set by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 7 days or at set failure (replace with another Quick-set). At day 28 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the X infusion set, entering Phase 2. All patients will be trained at this visit on the use of the X EW infusion set by site staff and demonstrate proficiency in the use of the EW infusion set. After 28 days or after using 4 sets, the patients will return all the extracted catheters sets .
  • Trial Arm B: Subjects will be randomized to a group who will be using the EWIS set infusion system for the initial Phase. All patients will be trained on the use of the EWIS infusion set by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 7 days or at set failure (replace with another Quick-set). At day 28 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the Quickset infusion set, entering Phase 2. All patients will then be retrained on the use of the Quickset infusion set by site staff and demonstrate proficiency. After 28 days or after using 4 sets, the patients will all the extracted catheters sets .

An extension phase for additional one month (4 sets) will be conducted as a single arm to assess the seven day durability of sets.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Clinical diagnosis of type 1 diabetes for at least one year
  • Using a MiniMed™ VEO pump or 640G Insulin pump with or without using an Enlite™ Sensor .
  • Age 18 to 80 years
  • Hemoglobin A1c level less than or equal to 10%
  • Not currently known to be pregnant, nor planning pregnancy during the study.
  • Willingness to follow the protocol and sign the informed consent

Exclusion criteria

  • A known medical condition that in the judgment of the investigator might interfere with the completion of the protocol.
  • Pregnant or lactating females

-

Treatment and study plan

7 day infusion set

Device

7 days survival of the sets

Primary outcomes

  1. Primary

    Time frame: 8 weeks

    • Percent of catheters reaching 4, 5, 6, 7 days use without set failure.

Secondary outcomes

  1. Exploratory outcomes

    Time frame: 8 weeks

    • Exploratory investigation of correlation between glycemia as measured by CGM records and pressure measurements, as recorded by the insulin pump.
    • Descriptive of site reactions. Pain at the infusion site device related issues (leakage, occlusion in fluid path, etc.).
    • Descriptive of set failure cases: accidental pull out, adhesive tape issues, infection, erythema and/or induration.
    • Descriptive of catheter explants:
    • Gross examination for cannula kinking and crimping.

Other outcomes

  1. Adverse Events

    Time frame: 8 weeks

    • Diabetic ketoacidosis (DKA)
    • Severe hypoglycemia
    • Severe hyperglycemia
    • Serious adverse events (SAEs)
    • Unanticipated adverse device effects (UADEs)

Sponsors and collaborators

Lead sponsor

Sheba Medical Center

Other Gov

Collaborators

  • Medtronic

Registry information

Important dates

Study start
2017
Primary completion
2020
Study completion
2020
First posted
Dec 23, 2019
Registry last updated
May 5, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.