NCT Number: NCT00747435

Safety and Effectiveness of the MED-EL Electric-Acoustic System

The purpose of this investigation is to demonstrate the safety and effectiveness of the MED-EL Electric-Acoustic System (EAS), a medical device that combines the use of a cochlear implant with an external electric-acoustic processor designed to provide benefit in speech perception and sound quality to individuals with a sensorineural hearing loss with minimal changes in residual hearing. The acoustic component of the processor will aid residual acoustic hearing in low frequency ranges, while the cochlear implant and electric component of the processor will be used to electrically stimulate the auditory nerve across a wide range of frequencies necessary for speech perception.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–70 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Stanford University Medical Center, Stanford, California, United States

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Audiologic tests suggest a moderate sloping to severe/profound sensorineural hearing loss in the ear to be implanted. The non-implanted ear may fall outside of these criteria; however, threshold levels cannot be better than the indication criteria (for example, the non-implanted ear must be 60 dB or poorer at 1000 Hz, 70 dB or poorer at 2000-8000 Hz).
  • Pure-tone air-conduction threshold levels for the ear to be implanted shall fall at or within the following levels:

250 Hz - 500 Hz hearing loss less than or equal to 65 dB HL 750 Hz hearing loss less than or equal to 75 dB HL 1000-1500 Hz hearing loss less than or equal to 60 dBHL (ie. 60-110+ dB HL) 2000 Hz - 8000 Hz hearing loss less than or equal to 70 dB HL

  • Pure-tone air-conduction threshold levels shall fall at or within the levels listed in the following chart Frequency (Hz): 250 500 750 1000 1500 2000 4000 8000 Lower Limit: 10 10 10 60 60 70 70 70 Upper Limit: 65 65 75 *110+ *110+ *110+ *110+ *90+
  • Pure-tone air-conduction thresholds for both ears are within 20 dB of each other at 250, 500 and 1000 Hz.
  • Air-bone gap at 500, 1000, 2000 and 4000 Hz should be <10 dB at two or more of these frequencies.
  • Minimal benefit from optimally fit hearing aid/s, preferably bilateral, with monosyllabic word scores in quiet of ≤50% in the best-aided condition.
  • Normal middle ear anatomy and function (based on clinical assessment of tympanometry and acoustic reflex results). No prior middle ear surgery or history of postadolescent, chronic middle ear infections or inner ear disorders (i.e. vertigo or Meniere's syndrome).
  • No evidence that hearing loss origin is retrocochlear.
  • Current user of bilateral acoustic hearing aids for at least 3 months. Note: If the subject has not been a successful hearing aid user (i.e. improper fit, feedback and/or discomfort from high frequency amplification), he/she must complete a hearing aid trial in the ear to be implanted for at least 1 month. If the ear to be implanted is poorer than the contralateral ear, then the subject must complete a hearing aid trial in both ears for at least 1 month to ensure that he/she is tested in the best-aided condition.
  • Adults 18-70 years of age at time of implantation.
  • Persons who currently exhibit sufficient understanding and communicative skills to comprehend general conversation in an "oral/aural mode" through normal conversation channels.
  • English as primary language.
  • Appropriate motivation and expectation levels.

Expanded Criteria:

Study Arm 2 candidates are those who meet the inclusion criteria listed above with the exception that:

In the ear to be implanted, the pure-tone air conduction threshold(s) at 250, 500, and/or 750 Hz will be greater than or equal to 0 dB HL and less than 10 dB HL, and/or In the ear to be implanted, the pure-tone air conduction threshold(s) at 1000 and/or 1500 Hz will be greater than or equal to 0 dB HL and less than 60 dB HL and/or In their best-aided condition, they score 51-60% on monosyllabic word scores.

Exclusion criteria

  • Conductive, retrocochlear or central auditory disorders.
  • Hearing loss in the ear to be implanted that has demonstrated a recent fluctuation at two or more frequencies of 15 dB in either direction in the last 2 years as demonstrated by serial audiograms. Note: In cases where 2-year documentation is missing, documentation of at least 18 months is necessary along with patient report of no fluctuation in the past 2 years. If 18-month data is unavailable, the patient must be monitored until 18-month data is available.
  • Any physical, psychological, or emotional disorder that would interfere with surgery or the ability to perform on test and rehabilitation procedures.
  • Developmental delays or organic brain dysfunction.
  • Physical or geographic limitations that may interfere with the completion of scheduled follow-up evaluations.
  • Skin or scalp conditions that could preclude magnetic attachment of the speech processor or use of the acoustic hearing aid.

Treatment and study plan

Electric Acoustic System

Device

Combination of a cochlear implant and a hearing aid

Other names: EAS, Electric Acoustic Systems

Primary outcomes

  1. Improvement in Speech Perception in Noise With EAS When Compared to the Preoperative Hearing Aid Alone Condition.

    Time frame: 12 months post initial activation

    CUNY sentences in noise are scored as the percent correct of words in each sentence. The total percent correct for preoperative hearing aid use was subtracted from the total percent correct for EAS at 12 months giving a percentage point improvement in speech perception with EAS.

Secondary outcomes

  1. Improvement in Speech Perception in Noise With EAS When Compared to the Cochlear Implant Alone Condition

    Time frame: 12 months initial activation

    CUNY sentences in noise are scored as the percent correct of words in each sentence. The total percent correct for the CI Alone condition was subtracted from the total percent correct for the EAS condition at 12 months giving an improvement in speech perception with EAS compared to CI Alone.

  2. Improvement on Speech Perception in the CI-only Condition as Compared to the Preoperative With Hearing Aid Condition.

    Time frame: 12 months post initial activation

    CNC words are scored as the percent correct out of 50 words. The percent correct for preoperative hearing aid use was subtracted from the percent correct for CI Alone at 12 months giving an improvement on speech perception for CI Alone compared to preoperative hearing aids.

  3. Increased Benefit With EAS as Compared to Their Preoperative Hearing Aid Condition as Measured by the APHAB Questionnaire.

    Time frame: 12 months post initial activation

    A lower score on the APHAB indicates a better performance. The global score for preoperative hearing aid use was subtracted from the global score for EAS at 12 months giving an improvement on the APHAB with EAS.

  4. Increase in Satisfaction With EAS Compared to Preoperative Hearing Aid Condition as Measured by the HDSS.

    Time frame: 12 months post initial activation

    Satisfaction is ranked from 1-5, with 1 being very dissatisfied and 5 being very satisfied. Data reported is the percentage of subjects who experienced an increase in satisfaction when using EAS compared to preoperative hearing aids.

Sponsors and collaborators

Lead sponsor

Med-El Corporation

Industry

Registry information

Official study title

The MED-EL EAS (Electric-Acoustic System) Using the PULSARCI100 FLEXeas / SONATATI100 FLEXeas and the DUET Speech Processor

Acronym: EAS

Important dates

Study start
2007
Primary completion
2016
Study completion
2016
First posted
Sep 5, 2008
Registry last updated
May 17, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.