Safety and Effectiveness of BioTraceIO Lite for Tissue Damage Assessment Following Liver Tissue Ablation Procedures

Clinical study planned to demonstrate that the BioTraceIO Lite, available post-procedure, is effective at estimating the area of tissue damage as measured on 24-hour post-procedure (T=24hrs) CECT scan and that it is safe, based on an assessment of device-related Adverse Events.

Multi-center (up to 6 investigational sites) prospective single-arm clinical investigation 50 evaluable subjects total stratified by cancer type (primary hepatocellular carcinoma vs. secondary liver metastases).

Completed

Looking for future studies?

Notify Me

Key information

Age range

21 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Mayo Clinic, Phoenix, Arizona, United States

Loading trial locations.

About this study

Clinical study planned to demonstrate that the BioTraceIO Lite, available post-procedure, is effective at estimating the area of tissue damage as measured on 24-hour post-procedure (T=24hrs) contrast-enhanced CT scan and that it is safe, based on an assessment of device-related Adverse Events.

BioTraceIO Lite is intended to provide physicians with adjunctive information in their clinical assessment of tissue damage created by liver tissue ablation, as part of their overall post-procedure clinical assessment.

Multi-center (up to 6 investigational sites) prospective single-arm clinical investigation 50 evaluable subjects total stratified by cancer type (primary hepatocellular carcinoma vs. secondary liver metastases).

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Scheduled and indicated for standard-of-care liver tissue ablation for either hepatocellular carcinoma (HCC) or metastatic liver tumor(s) using either RF or MW energy.
  • At least 21 years of age
  • Single tumor, or multiple tumors only if the distance between the ablated tumor and all other tumors allows for distinct separation between the necrotic zones (minimum of 1cm), based on the physician's discretion.
  • Distance between the tumor and the edge of any previous necrotic zones allows for distinct separation between the necrotic zones (minimum of 1cm), based on the physician's discretion.
  • Single ablation, using a single ablation needle, per tumor.
  • Able and willing to give informed consent.

Exclusion criteria

  • Liver tumor that cannot be ablated with a single ablation needle, according to the investigator's clinical opinion.
  • Subject cannot tolerate/undergo contrast-enhanced CT.
  • Planned ablation includes adjunctive means other than RF or MW energy (e.g., ethanol, hepatic artery embolization, etc.) or overlapping ablations using a single ablation needle.
  • Ablation area cannot be visualized continuously using ultrasound throughout the entire ablation procedure.
  • Pregnant or lactating
  • Currently participating in another clinical trial of an unapproved investigational device or drug that has not concluded the follow-up period.
  • Unable or unwilling to give informed consent.

Treatment and study plan

BioTraceIO Lite

Device

The BioTraceIO software will analyze the ultrasound images, and provide the physician with an estimated tissue damage map.

Primary outcomes

  1. Safety - Adverse Events

    Time frame: 1 year

    Incidence and severity of device-related adverse events (including serious adverse events (SAE)) occurring between the beginning of the liver tissue ablation procedure and the completion of the T=24hrs contrast-enhanced computed tomography (CECT) scan. There is no comparator for this endpoint.

  2. Effectiveness - DICE

    Time frame: 1 year

    Paired comparison of the DICE similarity coefficient, assessed between BioTraceIO Lite, as measured at end of ablation procedure (T=0) and the tissue damage area based on 24-hours post-procedure CECT (T=24), and the DICE similarity coefficient, assessed between the tissue damage area based on immediately post-procedure CECT (T=0) and the tissue damage area based on 24-hours post-procedure CECT (T=24).

Secondary outcomes

  1. Sensitivity

    Time frame: 1 year

    Sensitivity of the BioTraceIO Lite at T=0 compared to CECT at T=24hrs.

  2. Precision

    Time frame: 1 year

    Precision of the BioTraceIO Lite at T=0 compared to CECT at T=24hrs.

  3. Questionnaire - Effect of BioTraceIO

    Time frame: 1 year

    The proportion of subjects in whom the treating physician indicates that, if the BioTraceIO Lite results had been available for use in patient management, it would have impacted their follow-up plan for patient management, compared to the CECT alone.

  4. Difference between T=0 and T=24hrs CECT tissue damage

    Time frame: 1 year

    Difference in percentage between tissue damage volume as visualized at T=0 contrast-enhanced computed tomography (CECT) versus tissue damage volume as visualized at T=24hrs CECT.

Sponsors and collaborators

Lead sponsor

Techsomed Medical Technologies LTD

Industry

Registry information

Official study title

Pivotal Clinical Study to Establish the Safety and Effectiveness of the Use of BioTraceIO Lite for Assessment of Tissue Damage Following Liver Tissue Ablation Procedures

Important dates

Study start
2021
Primary completion
2022
Study completion
2022
First posted
Jul 21, 2021
Registry last updated
Apr 18, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.