Skip to main content
OpenTrials
Completed

NCT Number: NCT04435132

Robotic-assisted Percutaneous Access

This is a first-in-man (FIM) study to show feasibility and safety profile of the newly developed robotic device for percutaneous access in PCNL surgery. Patients with renal stone disease who had standard indications for prone PCNL were included in the study. PCNL was conducted with percutaneous puncture with the aid of robotic device with the use of intra-operative fluoroscopy. 5 patients were recruited in this pilot study over a period of 3 months. Punctures were performed by the board-qualified urologist.

Completed

Looking for future studies?

Notify Me

Key information

Age range

21 year–75 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University Malaya Medical Centre

Kuala Lumpur, 59100, Malaysia

About this study

All patients underwent preoperative investigations and imaging study (CT Urogram/Intravenous Pyelography) done prior to the operation as per standard operation preparation. Intraoperatively fluoroscopic imaging was done for the patients and the images were synchronized to the software. Robotic arm was placed on the site of surgery as for prone PCNL and diaphragm of the arm was centred over the site of desire calyx of entry. A needle was engaged into the centre of the diaphragm and using the software the need was aligned in the line of puncture automatically based on the fluoroscopic images. After the alignment was done the needle was advanced manually by the surgeon until the entry was confirmed by efflux of urine from the puncture needle. After the entry was confirmed a standard PCNL using 24 F or 30 F Balloon dilator was performed and stone cleared using Cyberwand/ Shockpulse etc. All the intraoperative and perioperative parameters as mentioned in the data collection form was measured. Results was analysed with the main focus on time of obtaining the puncture, accuracy of puncture, radiation exposure time till calyceal entry which was confirmed by free flow of urine and for any complications.

Patients aged between 21 to 75 years old who were diagnosed with kidney stone diseas and planned for prone PCNL with indication for surgery as per standard of care were recruited. They understood the purpose of testing and offered their voluntary and informed consent.

The robotic device i.e. ANT lesses the learning curve of percutaneous access as it reduces the need for surgeon's visualization of the calyceal system and the needle alignment. It also improved efficiency as there is less dependence on human skills during the procedure. As a whole, the FIM trial was a success. The performance of the ANT system during the trial was satisfactory. The robot alignment time is less than 30 seconds. Total puncture time was around 6 minutes and no AEs were reported.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age between 21 to 75 years
  • Diagnosed with kidney stone disease and planned for prone PCNL with indication for surgery as per standard of care
  • Subject who understand the purpose of testing, voluntary and informed consent, patients undergoing invasive imaging follow-up.

Exclusion criteria

  • Bleeding disorders
  • Patient on anticoagulation/antiplatelets drugs
  • Pyonephrosis
  • Pregnant women
  • Not able to have prone position for the procedure due to comorbidities
  • Moderate to severe renal failure
  • The patient's life expectancy is less than 12 months
  • Poor compliance and patients unable to complete the study in accordance with the requirements.

Treatment and study plan

PCL under fluoroscopy with the aid of the robotic device

Device

During the surgery, the robot was positioned over the estimated insertion point. It was aligned to be parallel to the fluoroscope imaging plane as much as possible. The needle is to be placed through the holder, and slightly penetrate the surface of the skin. An x-ray image is then taken to ensure that the whole device can be seen on the screen. Image calibration is then performed through the software GUI. After which, the clinician is also to perform needle tip selection. After which, needle alignment will then be performed. Once needle alignment is complete, the clinician will then advance the trocar needle into the patient's body, using the fluoroscopy to judge the penetration depth. Once the target has been reached, the needle style is to be extracted. The indication of leakage of fluid through the outer needle confirms the success of the procedure.

Primary outcomes

  1. Needle insertion time

    Time frame: During procedure

    Time taken for PCNL procedure

  2. Needle alignment time

    Time frame: During procedure

    Time taken for needle alignment

  3. Time to target

    Time frame: During procedure

    Time taken for the needle to reach the calyx, from on the skin to successful needle targeting

  4. Number of needle insertion attempts to obtain access

    Time frame: During procedure

    Number of attempts before successfully reaching the calyx

  5. Radiation dosage from placement of robot to efflux of urine

    Time frame: During procedure

    A measure of the radiation exposed during the surgical procedure

  6. Radiation dosage from placement of robot to successful needle target via image confirmation

    Time frame: During procedure

    A measure of the radiation exposed during the surgical procedure

Sponsors and collaborators

Lead sponsor

NDR Medical Technology Pte Ltd

Industry

Collaborators

  • University of Malaya

Registry information

Official study title

Robotic-assisted Percutaneous Access: Interventional, Non-randomized, Open-label, Non-comparative and First-in-Man Study

Important dates

Study start
2018
Primary completion
2018
Study completion
2018
First posted
Jun 17, 2020
Registry last updated
Jun 17, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.