NCT Number: NCT05246358

Respiratory Rate Validation Study - ChroniSense Polso

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–52 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Clinimark Desaturation Laboratory, Site ID# 001

Louisville, Colorado, 80027, United States

About this study

20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso.

A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.

Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Subject must have the ability to understand and provide written informed consent
  • Subject is adult over 18 years of age
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker, or a smoker who has refrained from smoking for 1 day.
  • Male or female of any race

Exclusion criteria

Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)

  • Subjects with known respiratory conditions such as:
  • uncontrolled / severe asthma,
  • flu,
  • pneumonia / bronchitis,
  • shortness of breath / respiratory distress,
  • respiratory or lung surgery,
  • emphysema, COPD, lung disease
  • Subjects with self-reported heart or cardiovascular conditions such as:
  • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg
  • have had cardiovascular surgery
  • chest pain (angina)
  • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia
  • previous heart attack
  • blocked artery
  • unexplained shortness of breath
  • congestive heart failure (CHF)
  • history of stroke
  • transient ischemic attack
  • carotid artery disease
  • myocardial ischemia
  • myocardial infarction
  • cardiomyopathy
  • Self-reported health conditions as identified in the Health Assessment Form
  • diabetes,
  • uncontrolled thyroid disease, kidney disease / chronic renal impairment,
  • history of seizures (except childhood febrile seizures),
  • epilepsy,
  • history of unexplained syncope,
  • recent history of frequent migraine headaches,
  • recent head injury within the last 2 months,
  • Cancer / chemotherapy
  • Other known health condition, should be considered upon disclosure in health assessment form

Treatment and study plan

ChoniSense Polso

Device

A range of stable respiratory rates was elicited from each volunteer test subject. The rates were 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.

Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.

Primary outcomes

  1. Arms accuracy (breaths per minute) of Polso Monitoring System relative to End Tidal Carbon Dioxide monitor-derived respiration rate

    Time frame: Through study completion, 1 month average

    Accuracy of Polso Monitoring System (DUT) relative to a reference End Tidal Carbon Dioxide monitor-derived respiration rate (REF). Polso reports respiration rate (breaths per minute) directly, at 60 second intervals. End Tidal Carbon Dioxide monitor waveforms will be hand-scored, based on respiration waveform peaks, to derive respiration rate (breaths per minute). The readings from each device will be synchronized to achieve paired simultaneous respiration rates, and aggregated into calculation of a comparative accuracy measure, namely Accuracy root-mean-square (Arms) difference, in breaths per minute, between the Polso (DUT) and the reference (REF) respiratory rate system, for all stable respiratory periods. Acceptance criteria Arms <3.

Sponsors and collaborators

Lead sponsor

ChroniSense Medical Ltd.

Industry

Registry information

Important dates

Study start
2021
Primary completion
2021
Study completion
2021
First posted
Feb 18, 2022
Registry last updated
Feb 18, 2022

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.