Clinimark Desaturation Laboratory, Site ID# 001
Louisville, Colorado, 80027, United States
NCT Number: NCT05246358
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso
Looking for future studies?
Notify Me18 year–52 year
All sexes
Interventional
Not applicable
Louisville, Colorado, 80027, United States
20 volunteer test subjects entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the ChroniSense Polso.
A range of stable respiratory rates were elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.
Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoo in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)
A range of stable respiratory rates was elicited from each volunteer test subject. The rates were 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with some natural variation from these exact numbers.
Each subject was instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor was used as the Accuracy Reference Device (Reference) in this study.
Time frame: Through study completion, 1 month average
Accuracy of Polso Monitoring System (DUT) relative to a reference End Tidal Carbon Dioxide monitor-derived respiration rate (REF). Polso reports respiration rate (breaths per minute) directly, at 60 second intervals. End Tidal Carbon Dioxide monitor waveforms will be hand-scored, based on respiration waveform peaks, to derive respiration rate (breaths per minute). The readings from each device will be synchronized to achieve paired simultaneous respiration rates, and aggregated into calculation of a comparative accuracy measure, namely Accuracy root-mean-square (Arms) difference, in breaths per minute, between the Polso (DUT) and the reference (REF) respiratory rate system, for all stable respiratory periods. Acceptance criteria Arms <3.
ChroniSense Medical Ltd.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07477834
Air Pollution Exposure, HRV
Abu Dhabi, United Arab Emirates
View Trial DetailsNCT06911541
Capnography, Respiratory Rate
Bergen, Norway
View Trial DetailsNCT06674629
Diaphragm Excursion, Diaphragm Thickness Fraction
Buenos Aires, Argentina
View Trial DetailsNCT05263791
Respiratory Rate
New Taipei City, Taiwan
View Trial Details