This study does not involve a therapeutic intervention as standard radiation therapy treatment will be prescribed. This study involves one additional 4DCT scan (i.e. the Real-Time Gated 4DCT scan) acquired immediately before or following the conventional 4DCT scan. This will take place on the day of the patient's treatment simulation, as per the current standard of care. The scanning sequence (i.e. conventional first versus gated first) will be randomised.
The Real-Time Gated 4DCT is anticipated to take longer than the conventional 4DCT scan, due to its gated (beam-pause) nature. However, upper limits for timing will be implemented in the software, and the scan aborted for highly erratic breathing traces that would not benefit from a Real-Time Gated 4DCT scan.
Interested in participating?
Request Info18 year and older
All sexes
Interventional
Not applicable
Blacktown, New South Wales, 2148, Australia
Location status: Recruiting
Four-dimensional computed tomography (4DCT) is considered the standard of care for modern high-precision stereotactic ablative body radiotherapy (SABR) techniques. Clinical 4DCT works by acquiring CT slices and breathing data synchronously as the patient moves through the rotating X-ray imager. The CT slices are sorted into 5-10 'phase images' (mid-exhale, peak-exhale, mid-inhale, peak-inhale, etc.) and stitched together to form a 3D+breathing induced tumour/organ motion or '4D' representation of the breathing anatomy. Current best practice in radiation therapy is for clinicians to perform manual delineation or 'contouring' of lung tumours on one or more 4DCT phase images to ensure proper coverage by the treatment beam despite breathing motion. In the presence of irregular breathing, mismatches arise between CT slices that are in the same breathing phase but imaged at different couch positions. As a result, the images produced suffer from various kinds of discontinuities including truncation, duplication or overlap. Yamamoto found that irregular breathing caused at least one error >4mm in 90% (45 out of 50) abdominal and thoracic patient scans. The main consequence of 4DCT image errors is that they can introduce tumour volume and position uncertainties as large as 30% between different 4D phase bins or different observers.
This can potentially reduce the likelihood of tumour control by up to one-third and lead to unnecessary irradiation of healthy lung tissue, contributing to major dose-limiting side effects like radiation pneumonitis, which is symptomatic in up to 30% of patients and fatal in 2%. The proposed solution is Real-Time Gated 4DCT using prospective regularity gating, which will be implemented for the first time on lung cancer patients.
The Real-Time Gated 4DCT method detects and then pauses the CT beam during irregular breathing events. We will analyse the patient's breathing pattern to prospectively gate acquisition in real-time. Real-time gated 4DCT ensures that the X-ray imager is switched on only when the breathing is regular; the scan gets paused otherwise. The benefits for the patient of gating a 4DCT scan have been demonstrated in previous studies:
The main culprit for imaging errors is irregular breathing: i.e. natural breathing variations, which are heightened for lung cancer patients but not accounted for by clinical scanners. A solution to this problem was proposed, which is to gate the CT scanner (i.e. pausing the beam) whenever a breathing condition occurs that is likely to produce an imaging artefact.
Research Question The primary endpoint is that the number of image artifacts using the Real-time gated scan will produce fewer artefacts than the conventional scan.
Secondary endpoints involve examining image quality and the impact on treatment planning with the benefits to the patient being:
This clinical trial is a first-in-humans pilot feasibility study and the aim is therefore to prove feasibility of Real-Time Gated 4DCT, with the aim of leading to an efficacy trial. As such, patient scans will be acquired across a broad range of patient breathing conditions to optimise the Real-Time Gated technique and to plan a larger, hypothesis-driven phase II clinical trial to follow. Automatic cut-offs for these erratic breathing traces will be built into the Real-Time Gated 4DCT software.
Rationale for Current Study In-silico studies suggest that the Real-Time Gated 4DCT approach, can reduce imaging errors by up to 50% whilst also reducing imaging dose by >20%. It has been estimated that the current rate of artifacts using conventional 4DCT is at least 60% and possibly up to 95%. A reduction to at most 35% (25% of images with fewer artifacts) would be clinically worthwhile.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Real-time gated four-dimensional cone beam computed tomography (4DCT) will be conducted at the simulation session. The Real-Time Gated 4DCT will be conducted while the patient moves through the rotating X-ray imager. This method detects and then pauses the CT beam during irregular breathing events.
Four-dimensional computed tomography (4DCT), the standard of care for modern high-precision stereotactic ablative body radiotherapy (SABR), will be conducted at the simulation session. It will run continuously while the patient moves through the rotating X-ray imager.
Time frame: 30 minutes
A change in the anticipated rate of image artifacts in images acquired from Conventional of 60% to 35% or less in images acquired from Real-Time Gated 4DCT.
Time frame: 1 minute at end of planning session
The radiation therapist conducting the radiation therapy study scan will complete a survey consisting of two yes/no questions after the planning session to assess patient tolerance (1. whether the patient finishes the scan without getting off the couch. 2. If the patient voiced any complaints during the scan and seem uncomfortable.)
Time frame: 5 minutes at end of planning session
The radiation therapist conducting the radiation therapy study scan will complete an in-house survey of 10 questions after the treatment session to evaluate their experience with REACT software and its usability.
Time frame: 5-10 minutes, following delineation of standard care and study scans.
The radiation oncologist performing the delineation on the radiation therapy planning scan images and study scan images will complete an in-house survey of 10 questions to evaluate his experience with REACT and his confidence with using REACT for delineation.
Time frame: 40 minutes, prior to and during the planning session
Correlation between image quality (the number of artifacts greater than 4mm in the images acquired from Real-Time Gated 4DCT) and participant characteristics known to affect image quality in conventional scans: Breathing period ( seconds) and breathing amplitude will be measured using the Varian real-time position management system; body mass index (weight(kg) x height(m)2); tumour location;, breathing regularity (root mean square error (RMSE) of 6 breaths/min was obtained), pulmonary function, Performance status (Eastern Cooperative Oncology Group (ECOG) performance scale).
Time frame: 30 minutes
Time in minutes from start to end of scan.
Time frame: 1 week
Contouring and planning using the Real-Time Gated 4DCT image set to develop a treatment plan of prescribed radiation therapy. Treatment plans will be assessed using the 3%/3mm and 2%/2mm gamma pass index.
Contact information is provided by the study sponsor or research team.
University of Sydney
Other
Respiratory Adaptive Computed Tomography: A Pilot Feasibility Study on the Use of Real-Time Gated 4DCT for Lung Cancer Radiation Therapy
Acronym: REACT
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06540196
Bronchial Neoplasms, Carcinoma, Bronchogenic
Hamilton, Ontario, Canada
View Trial DetailsNCT06410300
Lung Cancer, Lung Diseases
Columbia, Missouri, United States
View Trial DetailsNCT01441089
Adnexal Diseases, Breast Cancer
Bethesda, Maryland, United States
View Trial DetailsNCT00034216
Breast Cancer, Breast Diseases
Bethesda, Maryland, United States
View Trial Details