The Ohio State University Wexner Medical Center
Columbus, Ohio, 43210, United States
NCT Number: NCT01837537
The purpose of this study is to meet the respiration rate accuracy specifications for the study pulse oximeter monitoring system (study device) in a hospitalized volunteer population when compared to established technologies for measuring parameters related to respiration, such as Capnograph Respiration Rate (RR)(CO2).
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Columbus, Ohio, 43210, United States
Volunteer subjects from a hospital setting (see specific inclusion/exclusion criteria) over the age of 18, male and female, who provide written informed consent prior to the procedure, and are without any contact skin allergies to adhesives found in standard pulse oximetry sensors. A cross section of demographics - cardiovascular, respiratory, and metabolic type cases should be considered for a broad distribution of medical and surgical patients.
The study consists of measuring parameters related to respiration in hospitalized volunteers by means of the Bedside Respiratory Patient Monitoring System with RR V2.0 (Nightingale). The production-equivalent Nightingale device and multiparameter reference devices will be attached to the subject via sensors, nasal cannulas and ECG leads and the information from all devices will be captured continuously for up to 40 minutes to allow for at least 30 minutes of artifact free data 30 minutes.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via Adult Respiratory Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography).
Time frame: up to 40 minutes of continuous monitoring
The software calculates respiration rate values via the Max-N Sensor with a mean error of +/- 1 breath per minute relative to a capnography based reference. The Mean Error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the sensor and the Respiration Rate calculated from capnography). The error value estimates the mean difference in simultaneous estimates of respiration rate (the Respiration Rate calculated from the Max-N sensor minus the Respiration Rate calculated from capnography.
Medtronic - MITG
Industry
Evaluation of Respiration Rate Parameters in a Hospital Population Using a Nellcor Bedside Respiratory Patient Monitoring System With Respiration Rate Version 2.0
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07477834
Air Pollution Exposure, HRV
Abu Dhabi, United Arab Emirates
View Trial DetailsNCT06911541
Capnography, Respiratory Rate
Bergen, Norway
View Trial DetailsNCT06674629
Diaphragm Excursion, Diaphragm Thickness Fraction
Buenos Aires, Argentina
View Trial DetailsNCT05263791
Respiratory Rate
New Taipei City, Taiwan
View Trial Details