Singapore Eye Research Institute
Singapore, 168751
NCT Number: NCT01933165
Dry eye clinical research is limited to follow up of disease progression. While a large catalogue of tests, such as corneal staining grading, tear break up time (TBUT) and Schirmer's test, is available, objective assessments demonstrate poor to moderate repeatability. Among these assessments, TBUT has the best repeatability while the Schirmer's test shows good repeatability for more severe forms of the disease.1 As such, currently, most research outcomes are based on subjective patient complaints rather than objective assessment of the condition.
LipiView Ocular Surface Interferometer (LipiView) is a HSA-approved imaging device that has the ability to assess optical interference pattern produced by light reflected off the tear film in a safe and non-invasive manner.2 Software analysis of the image then measures the tear LLT, which is used to objectively quantify the diagnosis of lipid-deficient dry eye.
This current study aims to investigative the inter and intra-examiner reproducibility of measuring tear lipid layer thickness (LLT) with LipiView, which is used in our current clinical setting. LipiView has not been evaluated on its reproducibility of measuring tear lipid layer thickness. The study will also assess for correlation between the magnitude of tear LLT and the inter and intra-examiner reproducibility of LipiView.
Once this method is found to be repeatable, this can be used in future clinical settings and trials as an additional tool to investigate dry eye treatments.
Looking for future studies?
Notify Me21 year–90 year
All sexes
Interventional
Not applicable
Singapore, 168751
This study will be conducted in compliance with the protocol, SGCP and the applicable regulatory requirements.
Dry eye is a condition that affects 5-35% of the population.11 Common symptoms of this condition include eye irritation, heaviness of the eyelids, visual disturbances, tearing and light sensitivity.12 This significantly impacts the quality of life and vision of dry eye patients. In addition, there is a significant socioeconomic burden. In 2009, 54,051 patients sought treatment for dry eye in the Singapore National Eye Centre (SNEC). In total, the costs for dry eye medication amounted to $181,354.17.14 Additionally, dry eye affects work productivity which further adds on to the socioeconomic burden.15
Currently, the diagnosis and assessment of dry eye is hindered by tests that have poor repeatability, some of which cause discomfort due to their invasiveness. As a result most clinical studies involving dry eye treatment rely on symptomatic improvements as a primary outcome measure. However, symptomatic improvements are a poor measure in determining treatment effectiveness because with disease progression, the eye becomes desensitized from nerve damage. Also, it is common for patients to feel a reduction of discomfort of the eye before objective improvements can be noted by the clinician. An assessment based on symptoms alone is therefore inappropriate.13
Over the last few years, SNEC/SERI has built up a dedicated team of dry eye researchers for trials and evaluation of diagnostic devices. There is an increasing interest internationally, especially from industry, to partner us for testing of commercial products.
The LipiView Ocular Surface Interferometer is a HSA-approved imaging device that allows clinicians evaluate the characteristics and thickness of the tear lipid layer thickness (LLT) in a safe and non-invasive manner. The device utilizes white light from a LED source that is directed at the tear film over the inferior corneal region. Reflected light is captured by a high resolution camera and the optical inference pattern is evaluated by software to determine the LLT. 2 It also was found to correlate with other dry eye diagnostic tests, such as tear film break-up time (TBUT), Schirmer's test and corneal staining 3,4,5,6. LLT is also reported to better correlate with symptoms than other objective dry eye tests. 2
While several studies have investigated and shown the association between dry eye syndrome and low LLT 2,7,8,9, it is not known if LipiView would be able to make reproducible measurements of LLT for the objective diagnosis of dry eye syndrome.
Clinical importance Should this modality of imaging be shown to be repeatable, it can be incorporated into current protocol/workflow for assessment and monitoring of dry eye and treatment progress in our centre, as well as enhancing our capability to perform clinical trials and increase our international reputation.
Study Objectives and Purpose The primary purpose is to evaluate intra-examiner repeatability of the LipiView Ocular Surface Interferometer (LipiView) for imaging tear lipid layer thickness.
Study design:
Prospective study
Rationale:
Repeatability is an important issue in the objective assessment of dry eye. There is lack of studies that investigate LipiView for assessment of the tear film lipid layer thickness in dry eyes. LipiView is potentially useful as a non-invasive procedure for objective measurements of dry eyes.
Methods:
Participants and target sample size 20 healthy volunteers will be recruited for this study. During recruitment, a simple questionnaire is given to confirm for the absence of any dry eye symptoms. Each volunteer will also be given a study serial number to ensure anonymity.
All 20 volunteers will participate in the LipiView repeatability and inter-rater agreement study.
Visit schedules One visit is required for participants.
Duration of study:
10 months.
Procedures:
Similar steps would be taken for the other eye. Therefore, 12 measurements would be taken altogether per participant. Between measurements, participants are advised not to rub or touch their eyes, or instill any eye drops or ointment.
Assessment of fluorescein tear break up time and corneal staining will be conducted only after LipiView LLT measurement.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 1 day
Singapore National Eye Centre
Other Gov
Inter and Intra Examiner Reproducibility of LipiView Ocular Surface Interferometer (LipiView) for Measuring Tear Lipid Layer Thickness
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06071780
Dry Eye Syndromes, Dry Eyes
Kuala Lumpur, Malaysia
View Trial DetailsNCT01682460
Dry Eye Syndromes, Dry Eyes
Waterloo, Ontario, Canada
View Trial DetailsNCT01679015
Allergic Conjunctivitis, Conjunctival Diseases
Scottsdale, Arizona, United States
View Trial DetailsNCT01695668
Dry Eye Syndromes, Dry Eyes
View Trial Details